Purpose: We present the multifocal electroretinogram (mfERG) with a 7-hexagon array as an objective test of macular function that can be recorded in 14 s. We provide normal values and investigate its reproducibility and validity.

Methods: Healthy participants underwent mfERG testing according to International Society for Clinical Electrophysiology of Vision (ISCEV) standards using the Espion Profile/D310 multifocal ERG system (Diagnosys, LLC, Lowell, MA, USA). One standard recording of a 61-hexagon array and 2 repeated recordings of a custom 7-hexagon array were obtained.

Results: A total of 13 subjects (mean age 46.9 years) were included. The median response densities were 12.5 nV/deg2 in the center and 5.2 nV/deg2 in the periphery. Intereye correlations were strong in both the center (ρCenter = 0.821; p < 0.0001) and the periphery (ρPeriphery = 0.862; p < 0.0001). Intraeye correlations were even stronger: ρCenter = 0.904 with p < 0.0001 and ρPeriphery = 0.955 with p < 0.0001. Bland-Altman plots demonstrated an acceptable retest mean difference in both the center and periphery, and narrow limits of agreement. We found strong correlations of the center (ρCenter = 0.826; p < 0.0001) and periphery (ρPeriphery = 0.848; p < 0.0001), with recordings obtained by the 61-hexagon method.

Conclusions: The 7-hexagon mfERG provides reproducible results in agreement with results obtained according to the ISCEV standard.

Download full-text PDF

Source
http://dx.doi.org/10.1159/000475996DOI Listing

Publication Analysis

Top Keywords

7-hexagon array
8
center ρcenter
8
00001 periphery
8
periphery ρperiphery
8
00001
6
7-hexagon
4
7-hexagon multifocal
4
multifocal electroretinography
4
electroretinography objective
4
objective functional
4

Similar Publications

Revealing a retinal facilitatory effect with the multifocal ERG.

Doc Ophthalmol

April 2019

Department of Ophthalmology, Faculty of Medicine, Research Institute of the McGill University Health Centre, 1001 Decarie BOULEVARD, Glen Site, Block E, Room EM03238, Montreal, QC, H4A 3J1, Canada.

Purpose: We have previously shown that the amplitude of the mfERG response obtained to a single (large) hexagon is significantly smaller than that obtained when summating all the mfERG responses evoked to an array of 7-61 hexagons covering the same retinal area. The purpose of this study was to confirm our initial findings in normal subjects of different ages and in selected patients.

Methods: Binocular mfERGs (1, 7, 19, 37 and 61 hexagon arrays; Espion V6.

View Article and Find Full Text PDF

Purpose: We present the multifocal electroretinogram (mfERG) with a 7-hexagon array as an objective test of macular function that can be recorded in 14 s. We provide normal values and investigate its reproducibility and validity.

Methods: Healthy participants underwent mfERG testing according to International Society for Clinical Electrophysiology of Vision (ISCEV) standards using the Espion Profile/D310 multifocal ERG system (Diagnosys, LLC, Lowell, MA, USA).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!