Purpose: To evaluate the in-scan and scan-rescan consistency of left ventricular (LV) in- and outflow assessment from 1) 2D planimetry; 2) 4D flow magnetic resonance imaging (MRI) with retrospective valve tracking, and 3) 4D flow MRI with particle tracing.

Materials And Methods: Ten healthy volunteers (age 27 ± 3 years) underwent multislice cine short-axis planimetry and whole-heart 4D flow MRI on a 3T MRI scanner twice with repositioning between the scans. LV in- and outflow was compared from 1) 2D planimetry; 2) 4D flow MRI with retrospective valve tracking over the mitral valve (MV) and aortic valve (AV), and 3) 4D flow MRI with particle tracing through forward and backward integration of velocity data.

Results: In-scan consistency between MV and AV flow volumes is excellent for both 4D flow MRI methods with r ≥ 0.95 (P ≤ 0.001). In-scan AV and MV flow by retrospective valve tracking shows good to excellent correlations versus AV and MV flow by particle tracing (r ≥ 0.81, P ≤ 0.004). Scan-rescan SV assessment by 2D planimetry shows excellent reproducibility (intraclass correlation [ICC] = 0.98, P < 0.001, coefficient of variation [CV] = 7%). Scan-rescan MV and AV flow volume assessment by retrospective valve tracking shows strong reproducibility (ICCs ≥ 0.89, P ≤ 0.05, CVs = 12%), as well as by forward and backward particle tracing (ICCs ≥ 0.90, P ≤ 0.001, CVs ≤ 11%). Multicomponent particle tracing shows good scan-rescan reproducibility (ICCs ≥ 0.81, P ≤ 0.007, CVs ≤ 16%).

Conclusion: LV in- and outflow assessment by 2D planimetry and 4D flow MRI with retrospective valve tracking and particle tracing show good in-scan consistency and strong scan-rescan reproducibility, which indicates that both 4D flow MRI methods are reliable and can be used clinically.

Level Of Evidence: 2 Technical Efficacy Stage: 2 J. Magn. Reson. Imaging 2018;47:511-522.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5811894PMC
http://dx.doi.org/10.1002/jmri.25792DOI Listing

Publication Analysis

Top Keywords

flow mri
32
retrospective valve
20
valve tracking
20
particle tracing
20
in- outflow
16
flow
13
assessment planimetry
12
planimetry flow
12
mri retrospective
12
iccs ≥
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!