Spike or not a spike? That is the question in a patient with single lead pacemaker.

J Electrocardiol

Servicio de Cardiología, Hospital Universitario de La Princesa, Madrid, Spain; School of Medicine, Universidad Autonoma, Madrid, Spain.

Published: August 2018

Dual-chamber pacing is feasible via the floating atrial sensor electrodes of a single-pass VDD lead but the atrial lead threshold is higher than the accepted clinical standard. Furthermore, due to the floating nature of the system, atrial sensing and pacing thresholds may vary during the follow up. For these reasons this strategy is seldom considered a common pacing solution in routine clinical practice. Alternatively, this phenomenon is likely to be observed as a result of incorrect generator configuration. As shown in our case, this inadequate setting can be suspected just by the analysis of the surface ECG and the post implantation chest X-ray.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jelectrocard.2017.06.018DOI Listing

Publication Analysis

Top Keywords

spike spike?
4
spike? question
4
question patient
4
patient single
4
single lead
4
lead pacemaker
4
pacemaker dual-chamber
4
dual-chamber pacing
4
pacing feasible
4
feasible floating
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!