Purpose: The purpose of this study was to describe and evaluate methods for calculating a megavoltage computed tomography (MVCT)-derived MR hardware attenuation map (μ-map) and dual-energy CT (DECT) for 511 keV photons.
Methods: Phantom measurements were acquired on a whole-body hybrid PET/MRI system, using a four-channel receive-only MR radiofrequency (RF) breast coil. Two acquisitions were performed: with the phantoms positioned in the four-channel RF breast coil, and without the breast coil. PET attenuation from the breast coil was corrected using three different CT-derived hardware μ-maps: (a) Single-energy CT (SECT), (b) DECT, and (c) MVCT. Each attenuation-corrected (AC) PET volume was evaluated and compared with the acquisition performed without the breast coil.
Results: The breast coil attenuated PET photons by 10% overall. Threshold values were applied to the SECT μ-map to reduce the effects of metal artifacts, but overcorrection occurred in more highly attenuated regions. The DECT-derived virtual monochromatic image reduced beam-hardening artifacts, but other metal artifacts remained. Despite the remaining metal artifacts in the DECT image, it led to an improvement in the more attenuated regions. The MVCT images appear to be free from metal artifacts leading to an artifact-free μ-map and a further improvement AC-PET images.
Conclusions: Our MVCT-based approach for creating μ-maps for MR RF coils greatly reduces artifacts produced by metal in a SECT approach. This eliminates the need for other artifact reduction methods, including the application of a threshold of narrow beam attenuation coefficients, or disassembling hardware to remove high-Z components before imaging with a kilovoltage source.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.12407 | DOI Listing |
Cell Biosci
December 2024
Department of Biological and Environmental Sciences and Technologies (DiSTeBA), University of Salento, Lecce, Italy.
N-myc downstream-regulated gene 1 (NDRG1) is a member of the NDRG family of intracellular proteins and plays a central role in a wide range of biological processes including stress response, differentiation, and metabolism. The overexpression of NDRG1 is an indicator of poor prognosis in various types of cancer. Here, we found that NDRG1 is an independent prognostic marker of poor outcome in breast cancer (BC).
View Article and Find Full Text PDFFront Bioeng Biotechnol
December 2024
Vascularized Composite Allotransplantation Laboratory, Massachusetts General Hospital, Harvard Medical School, Boston, MA, United States.
Introduction: Magnetic nanoparticles (MNPs), particularly iron oxide nanoparticles (IONPs), are renowned for their superparamagnetic behavior, allowing precise control under external magnetic fields. This characteristic makes them ideal for biomedical applications, including diagnostics and drug delivery. Superparamagnetic IONPs, which exhibit magnetization only in the presence of an external field, can be functionalized with ligands for targeted affinity diagnostics.
View Article and Find Full Text PDFProtein Sci
January 2025
Department of Biology, Texas Christian University, Fort Worth, Texas, USA.
Inherited mutations in the genes coding for the tumor suppressor proteins BRCA1 and PALB2 can lead to increased risk of breast and ovarian cancer. Upon DNA damage, these two proteins form a complex to promote double-stranded break repair via homologous recombination. Missense mutations in either BRCA1 or PALB2 that disrupt this important interaction result in loss of effective DNA damage repair and are associated with breast tumorigenesis.
View Article and Find Full Text PDFEur Radiol Exp
November 2024
Department of Radiology, Medical University Innsbruck, Innsbruck, Austria.
Background: To investigate the artifact sizes of four common breast clip-markers on a standard breast magnetic resonance imaging (MRI) protocol in an in vitro phantom model.
Methods: Using 1.5-T and 3-T whole-body scanners with an 18-channel breast coil, artifact dimensions of four breast biopsy markers in an agarose-gel phantom were measured by two readers on images obtained with the following sequences: T2-weighted fast spin-echo short inversion time fat-suppressed inversion-recovery with magnitude reconstruction (T2-TIRM); T1-weighted spoiled gradient-echo with fat suppression (T1_FL3D), routinely used for dynamic contrast-enhanced imaging; diffusion-weighted imaging (DWI), including a readout segmented echo-planar imaging (RESOLVE-DWI) and echo-planar imaging sequence (EPI-DWI).
Int J Numer Method Biomed Eng
December 2024
Electrical and Electronics Engineering Department, Middle East Technical University, Ankara, Türkiye.
This study investigates the pulse compression technique to improve the performance of magneto-acousto-electrical tomography (MAET) with magnetic field measurements through numerical studies. Emphasizing the effects of specific coil configuration on MAET measurements, the study conducts evaluations using a linear phased array (LPA) transducer and numerical breast models with tumor inclusion. It provides feasibility and a detailed comparative analysis of various excitations, including linear frequency modulated (LFM), Barker code, and Golay code excitations in MAET.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!