Aim: Critical illness and medical interventions, such as renal replacement therapy, can cause changes to vancomycin pharmacokinetics and lead to suboptimal dosing. To comprehensively characterize vancomycin pharmacokinetic a method must measure vancomycin in a range of clinical matrices.
Results: A LC-MS/MS method was developed using hydrophilic interaction liquid chromatography and microsample volumes, where possible. For all matrices, the linear concentration range was 1-100 μg/ml, interassay accuracy and precision was within 15%, and recovery above 80%. No matrix effects were observed. Calibration equivalence may be applied for some matrix combinations.
Conclusion: A method for the analysis of vancomycin in plasma (total, unbound), urine and renal replacement therapy effluent, suitable for use in any patient pharmacokinetic study, has been developed and validated.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4155/bio-2017-0019 | DOI Listing |