Purpose: To evaluate the repeatability and reproducibility of retinal blood flow (RBF) measurements in humans by using new auto-alignment and measurement software in a commercially available Doppler optical coherence tomography (DOCT) system.
Methods: The DOCT flowmeter assessed the intrasession repeatability and the intersession and interobserver reproducibility of the RBF measurements. For intrasession repeatability, the coefficients of variation (CVs) of five repeated RBF measurements were calculated at the retinal arteries and veins in 20 normal eyes of 20 healthy volunteers. For intersession reproducibility, two sets of three measurements obtained by one observer on 2 different days were compared. For interobserver reproducibility, two sets of three measurements obtained by two observers on the same day were compared. Intraclass correlation coefficients (ICCs) also were used to evaluate the repeatability and reproducibility. The relevance of the DOCT flowmeter and laser Doppler velocimetry (LDV) also was assessed.
Results: Regarding intrasession repeatability, the ICC of the RBF exceeded 0.90 in arterioles and venules (ICC: 0.994 and 0.970, respectively). The CVs of the RBF in the arterioles and venules were 6.0% ± 3.4% and 8.8% ± 5.1%, respectively. The intersession and interobserver RBF values had high reproducibility in the arterioles (ICC: 0.980 and 0.993, respectively) and venules (ICC: 0.982 and 0.986, respectively). The RBF measured with the DOCT flowmeter was correlated strongly with LDV in the arterioles (r = 0.76; P < 0.001).
Conclusions: The DOCT flowmeter had good reproducibility in the arterioles and venules and precisely measured the RBF as compared to the LDV in the arterioles.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1167/iovs.16-21389 | DOI Listing |
Transl Vis Sci Technol
November 2021
Department of Ophthalmology, Asahikawa Medical University, Asahikawa, Japan.
Purpose: We aimed to explore the velocity waveform characteristics of the retinal artery associated with age and the cardio-ankle vascular index (CAVI) as a conventional arterial stiffness marker by applying the Doppler optical coherence tomography (DOCT) flowmeter.
Methods: In this cross-sectional study, DOCT flowmeter imaging was performed in 66 participants aged 21 to 83 years (17 men, 49 women) with no history of eye diseases and no systemic diseases, except for hypertension. Retinal blood velocity waveform was analyzed where several parameters in time (upstroke time, T1, T2, T3, and T4) and area under the waveform (area elevation, area declination, A1, A2, A3, and A4) were extracted.
J Clin Med
June 2020
Department of Ophthalmology, School of Medicine, Asahikawa Medical University, Asahikawa 078-8510, Japan.
Background: Doppler optical coherence tomography (DOCT) flowmeter can be used to measure retinal blood flow (RBF) parameters, including vessel diameter, blood velocity, and the absolute value of RBF within 2.0 s. We investigated the RBF parameters in eyes with branch retinal vein occlusion (BRVO) using a DOCT flowmeter.
View Article and Find Full Text PDFPurpose: We aimed to observe longitudinal changes in retinal blood flow (RBF) and structural transformations in capillaries using Doppler optical coherence tomography (DOCT) and optical coherence tomography angiography (OCTA) in a feline retinal blood occlusion (RVO) model.
Methods: RVO was induced by argon green laser photocoagulation (PC) in six eyes of six cats. RBF was measured at a first-order retinal artery and vein by a DOCT flowmeter, and structural changes in the capillaries around the occluded vessels (12 × 12 and 3 × 3 mm) were assessed by OCTA before (at baseline); immediately after PC; and on days 1, 4, 7, and 14 thereafter.
Transl Vis Sci Technol
May 2019
Research and Development Division, Topcon Corporation, Tokyo, Japan.
Purpose: To evaluate the repeatability and reproducibility of the measurement of retinal arteriole pulse waveforms using a novel fully-automated Doppler optical coherence tomography (DOCT) flowmeter in healthy subjects.
Methods: Twenty eyes of 20 healthy subjects were included to test the intrasession repeatability of pulse waveform analysis. DOCT measurements were performed based on a newly developed instantaneous Doppler angle measurement method.
PLoS One
April 2019
Department of Ophthalmology, Asahikawa Medical University, Asahikawa, Japan.
To use a Doppler optical coherence tomography (DOCT) flowmeter to investigate segmental retinal blood flow (RBF) and sum of the segmental RBFs (SRBF) changes after panretinal photocoagulation (PRP) was used to treat type 2 diabetes mellitus with severe diabetic retinopathy (DR). Data from five patients with proliferative DR (PDR) (mean age 51.9 ± 10.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!