A novel approach to high-throughput, targeted liquid chromatography-tandem mass spectrometry (LC-MS/MS) analysis has been developed. A single chromatographic system can be used for the analysis of a range of 20 drugs and metabolites with a total analysis time of 36 s (one 96-well plate of prepared samples per hour). To demonstrate the applicability of this approach to quantitative analysis, a method has been validated for the therapeutic drug monitoring of clozapine and norclozapine following automated extraction from human plasma. Chromatographic retention times were 11.4 and 12.4 s for norclozapine and clozapine, respectively (for both analytes the chromatographic peak width was less than 1 s). Comparison with a conventional LC-MS/MS method (5 min analysis time) showed excellent agreement. This new approach offers analysis times more akin to flow-injection analysis, but is likely to be more widely applicable because of chromatographic resolution from residual matrix components and isobaric interferences.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/dta.2223 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!