Objectives: Free light chain (FLC) measurement gained a lot of interest for diagnostic workup of monoclonal gammopathy.

Methods: We evaluated the performance of turbidimetric polyclonal Freelite (The Binding Site, Birmingham, UK) assays on Cobas 6000 (Roche Diagnostics, Rotkreuz, Switzerland) and nephelometric monoclonal N Latex (Siemens Healthcare Diagnostics, Marburg, Germany) assays on BN ProSpec (Dade Behring, Deerfield, IL) vs established nephelometric Freelite assays on BN ProSpec.

Results: Analytical performance was acceptable. Method comparison (n = 118) showed significant proportional FLC differences for N Latex assays. However, good correlation and clinical concordance were shown. Recovery study in the low concentration range demonstrated consistent over- and underrecovery for Freelite reagents, hampering future research on prognostic value of suppressed noninvolved FLC. Antigen excess detection was successful for κ FLC in three-fourths of cases with Freelite reagents and in all cases with N Latex reagents. However, the latter resulted in underestimated κ FLC concentrations.

Conclusions: FLC analysis requires continuous awareness of analytical limitations. Monitoring of disease response requires FLC analysis on the same platform using the same reagents.

Download full-text PDF

Source
http://dx.doi.org/10.1093/ajcp/aqx037DOI Listing

Publication Analysis

Top Keywords

free light
8
light chain
8
freelite reagents
8
flc analysis
8
flc
7
reagents
5
performance evaluation
4
evaluation serum
4
serum free
4
chain analysis
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!