AI Article Synopsis

  • * Researchers analyzed eye movements in 23 awake patients with ICP monitored, finding that as ICP increased, eye-tracking metrics declined, particularly those related to cranial nerve VI.
  • * The findings suggest that eye tracking could serve as a noninvasive way to assess the impact of elevated ICP, aiding in diagnosing conditions like shunt malfunction and concussion.

Article Abstract

OBJECTIVE The precise threshold differentiating normal and elevated intracranial pressure (ICP) is variable among individuals. In the context of several pathophysiological conditions, elevated ICP leads to abnormalities in global cerebral functioning and impacts the function of cranial nerves (CNs), either or both of which may contribute to ocular dysmotility. The purpose of this study was to assess the impact of elevated ICP on eye-tracking performed while patients were watching a short film clip. METHODS Awake patients requiring placement of an ICP monitor for clinical purposes underwent eye tracking while watching a 220-second continuously playing video moving around the perimeter of a viewing monitor. Pupil position was recorded at 500 Hz and metrics associated with each eye individually and both eyes together were calculated. Linear regression with generalized estimating equations was performed to test the association of eye-tracking metrics with changes in ICP. RESULTS Eye tracking was performed at ICP levels ranging from -3 to 30 mm Hg in 23 patients (12 women, 11 men, mean age 46.8 years) on 55 separate occasions. Eye-tracking measures correlating with CN function linearly decreased with increasing ICP (p < 0.001). Measures for CN VI were most prominently affected. The area under the curve (AUC) for eye-tracking metrics to discriminate between ICP < 12 and ≥ 12 mm Hg was 0.798. To discriminate an ICP < 15 from ≥ 15 mm Hg the AUC was 0.833, and to discriminate ICP < 20 from ≥ 20 mm Hg the AUC was 0.889. CONCLUSIONS Increasingly elevated ICP was associated with increasingly abnormal eye tracking detected while patients were watching a short film clip. These results suggest that eye tracking may be used as a noninvasive, automatable means to quantitate the physiological impact of elevated ICP, which has clinical application for assessment of shunt malfunction, pseudotumor cerebri, concussion, and prevention of second-impact syndrome.

Download full-text PDF

Source
http://dx.doi.org/10.3171/2016.12.JNS161265DOI Listing

Publication Analysis

Top Keywords

elevated icp
16
eye tracking
16
film clip
12
icp
12
discriminate icp
12
icp ≥
12
elevated intracranial
8
intracranial pressure
8
impact elevated
8
patients watching
8

Similar Publications