  • The study aimed to evaluate the accuracy of cutaneous sensors compared to transosseous sensors for measuring rotational knee instability during the pivot shift test using cadaveric knees.
  • Eight cadaveric knees were tested for both intact and ACL-deficient conditions, measuring acceleration and angular velocity with sensors attached to the tibia.
  • Results showed that while cutaneous sensors correlated well with transosseous sensors, they had significantly lower measurements for overall acceleration and maximum internal rotation, suggesting these sensors may have limitations in clinical settings.

Article Abstract

Background: Recently several authors have reported on the quantitative evaluation of the pivot-shift test using cutaneous fixation of inertial sensors. Before utilizing this sensor for clinical studies, it is necessary to evaluate the accuracy of cutaneous sensor in assessing rotational knee instability. To evaluate the accuracy of inertial sensors, we compared cutaneous and transosseous sensors in the quantitative assessment of rotational knee instability in a cadaveric setting, in order to demonstrate their clinical applicability.

Methods: Eight freshly frozen human cadaveric knees were used in this study. Inertial sensors were fixed on the tibial tuberosity and directly fixed to the distal tibia bone. A single examiner performed the pivot shift test from flexion to extension on the intact knees and ACL deficient knees. The peak overall magnitude of acceleration and the maximum rotational angular velocity in the tibial superoinferior axis was repeatedly measured with the inertial sensor during the pivot shift test. Correlations between cutaneous and transosseous inertial sensors were evaluated, as well as statistical analysis for differences between ACL intact and ACL deficient knees.

Results: Acceleration and angular velocity measured with the cutaneous sensor demonstrated a strong positive correlation with the transosseous sensor (r = 0.86 and r = 0.83). Comparison between cutaneous and transosseous sensor indicated significant difference for the peak overall magnitude of acceleration (cutaneous: 10.3 ± 5.2 m/s, transosseous: 14.3 ± 7.6 m/s, P < 0.01) and for the maximum internal rotation angular velocity (cutaneous: 189.5 ± 99.6 deg/s, transosseous: 225.1 ± 103.3 deg/s, P < 0.05), but no significant difference for the maximum external rotation angular velocity (cutaneous: 176.1 ± 87.3 deg/s, transosseous: 195.9 ± 106.2 deg/s, N.S).

Conclusions: There is a positive correlation between cutaneous and transosseous inertial sensors. Therefore, this study indicated that the cutaneous inertial sensors could be used clinically for quantifying rotational knee instability, irrespective of the location of utilization.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jos.2017.04.010DOI Listing

Publication Analysis

Top Keywords

inertial sensors
20
cutaneous transosseous
16
quantitative evaluation
8
cutaneous
8
transosseous inertial
8
evaluate accuracy
8
cutaneous sensor
8
rotational knee
8
knee instability
8
pivot shift
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!