Accuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients under General Anesthesia in the Operating Room.

Anesthesiology

From the Department of Laboratory Medicine and Pathology (B.S.K., L.J.D.), Mayo Clinic School of Health Sciences (B.S.K., C.M.L., L.K.S., A.E.W., M.E.S.-M.), Department of Anesthesiology and Perioperative Medicine (C.M.L., L.K.S., A.E.W., M.E.S.-M., T.B.C.), Division of Biomedical Statistics and Informatics (C.M.W.-W.), and Department of Biomedical Engineering and Physiology (T.B.C.), Mayo Clinic, Rochester, Minnesota.

Published: September 2017

AI Article Synopsis

Article Abstract

Background: The aim of this study was to evaluate the use of a glucose meter with surgical patients under general anesthesia in the operating room.

Methods: Glucose measurements were performed intraoperatively on 368 paired capillary and arterial whole blood samples using a Nova StatStrip (Nova Biomedical, USA) glucose meter and compared with 368 reference arterial whole blood glucose measurements by blood gas analyzer in 196 patients. Primary outcomes were median bias (meter minus reference), percentage of glucose meter samples meeting accuracy criteria for subcutaneous insulin dosing as defined by Parkes error grid analysis for type 1 diabetes mellitus, and accuracy criteria for intravenous insulin infusion as defined by Clinical and Laboratory Standards Institute guidelines. Time under anesthesia, patient position, diabetes status, and other variables were studied to determine whether any affected glucose meter bias.

Results: Median bias (interquartile range) was -4 mg/dl (-9 to 0 mg/dl), which did not differ from median arterial meter bias of -5 mg/dl (-9 to -1 mg/dl; P = 0.32). All of the capillary and arterial glucose meter values met acceptability criteria for subcutaneous insulin dosing, whereas only 89% (327 of 368) of capillary and 93% (344 of 368) arterial glucose meter values met accuracy criteria for intravenous insulin infusion. Time, patient position, and diabetes status were not associated with meter bias.

Conclusions: Capillary and arterial blood glucose measured using the glucose meter are acceptable for intraoperative subcutaneous insulin dosing. Whole blood glucose on the meter did not meet accuracy guidelines established specifically for more intensive (e.g., intravenous insulin) glycemic control in the acute care environment.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5561475PMC
http://dx.doi.org/10.1097/ALN.0000000000001708DOI Listing

Publication Analysis

Top Keywords

glucose meter
36
capillary arterial
16
arterial blood
16
blood glucose
16
glucose
13
glucose measurements
12
meter
12
accuracy criteria
12
subcutaneous insulin
12
insulin dosing
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!