Impact of cone-beam computed tomography scan mode on the diagnostic yield of chemically simulated external root resorption.

Am J Orthod Dentofacial Orthop

Departments of Oral Pathology, Radiology and Medicine; Radiology; and Anatomy and Cell Biology, Colleges of Dentistry and Medicine, University of Iowa, Iowa City, Iowa.

Published: June 2017

Introduction: The aim of this in-vitro study was to evaluate the influence of cone-beam computed tomography scans on the diagnosis of chemically simulated external root resorption.

Methods: One hundred extracted anterior teeth were selected. Subsurface demineralization was induced on a limited area of the apical third of the root of 49 teeth. Each tooth was placed in an empty socket of a partially edentulous dry mandible. Cone-beam computed tomography images were obtained according to 3 protocols: (1) half scan, 0.40-mm voxel size; (2) full scan, 0.40-mm voxel size; and (3) full scan, 0.125-mm voxel size. Three observers evaluated the images. Sensitivity, specificity, accuracy, and area under the curve were compared with the Cochran Q and Mann-Whitney U tests.

Results: Protocol 3 had the highest sensitivity (81.63%), accuracy (80.67%), and area under the curve (0.807). There were statistically significant differences between protocol 3 and the other 2 protocols (P <0.001). The specificity of protocol 1 (84.97%) was greater than that of protocols 2 (69.93%) and 3 (79.74%); however, a statistically significant difference was found only between protocols 1 and 2 (P = 0.005).

Conclusions: A more dedicated, high-resolution scan should be acquired when one intends to investigate the early stage of external root resorption during orthodontic treatment. However, this does not imply that all orthodontic patients should be subjected to high-dose cone-beam computed tomography scans.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajodo.2016.10.041DOI Listing

Publication Analysis

Top Keywords

cone-beam computed
12
computed tomography
12
voxel size
12
chemically simulated
8
simulated external
8
external root
8
scan 040-mm
8
040-mm voxel
8
size full
8
full scan
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!