Millions of physically active individuals worldwide use heart rate monitors (HRMs) to control their exercise intensity. In many cases, the HRM indicates an unusually high heart rate (HR) or even arrhythmias during training. Unfortunately, studies assessing the reliability of these devices to help control HR disturbances during exercise do not exist. We examined 142 regularly training endurance runners and cyclists, aged 18-51 years, with unexplained HR abnormalities indicated by various HRMs to assess the utility of HRMs in diagnosing exertion-induced arrhythmias. Each athlete simultaneously wore a Holter electrocardiogram (ECG) recorder and an HRM during typical endurance training in which they had previously detected "arrhythmias" to verify the diagnosis. Average HRs during exercise were precisely recorded by all types of HRMs. No signs of arrhythmia were detected during exercise in approximately 39% of athletes, and concordant HRs were recorded by the HRMs and Holter ECG. HRMs indicated surprisingly high short-term HRs in 45% of athletes that were not detected by the Holter ECG and were artifacts. In 15% of athletes, single ventricular/supraventricular beats were detected by the Holter ECG but not by the HRM. We detected a serious tachyarrhythmia in the HRM and Holter ECG data with concomitant clinical symptoms in only one athlete, who was forced to cease exercising. We conclude that the HRM is not a suitable tool for monitoring heart arrhythmias in athletes and propose an algorithm to exclude the suspicion of exercise-induced arrhythmia detected by HRMs in asymptomatic, physically active individuals.

Download full-text PDF

Source
http://dx.doi.org/10.1111/sms.12917DOI Listing

Publication Analysis

Top Keywords

holter ecg
16
heart rate
12
rate monitors
8
physically active
8
active individuals
8
arrhythmia detected
8
detected holter
8
hrms
7
detected
6
hrm
5

Similar Publications

Electrical cardioversion presents one of the treatment options for atrial fibrillation (AF). However, the early recurrence rate is high, reaching ~40% three months after the procedure. Features based on vectorcardiographic signals were explored to find association with early recurrence of AF.

View Article and Find Full Text PDF

We aimed to assess the uric acid-to-high-density lipoprotein cholesterol (HDL-C) ratio (UHR) and several other parameters with respect to their performance in detecting recurrence among patients with atrial fibrillation (AF) who underwent ablation. This retrospective cohort study analyzed data from patients who underwent radiofrequency or cryoablation for paroxysmal, persistent, or long persistent AF between September 2021 and September 2023. After ablation, patients were monitored for 24 h, with an ECG Holter used for symptomatic cases.

View Article and Find Full Text PDF

Objectives: The objective of this study was to assess the quality of ECG recordings and the concordance between the automatic detection of cardiac arrhythmia episodes by a patch ECG and an insertable cardiac monitor.

Design: Prospective cohort study.

Setting And Participants: Endurance athletes diagnosed with paroxysmal atrial fibrillation (AF) and no other relevant comorbidities participating in a randomised controlled trial on the effects of training adaption.

View Article and Find Full Text PDF

Background: Atrial fibrillation (AF) is the most common cardiac arrhythmia. Given its often-paroxysmal nature, screening at a single time point, using a 12-lead electrocardiogram (ECG) or a Holter monitor, has limited benefit. The AliveCor KardiaMobile device is a validated ECG recorder that can be used for patient-directed arrhythmia diagnosis and symptom-rhythm correlation.

View Article and Find Full Text PDF

Background: Numerous studies have demonstrated the significance of trimethylamine-N-oxide (TMAO) in the progression of atrial fibrillation (AF). However, the association between TMAO and AF recurrence (RAF) post-catheter ablation is not yet fully understood. This study aims to elucidate the predictive capability of pre-procedural TMAO levels in determining RAF following catheter ablation (CA).

View Article and Find Full Text PDF