Guiding flying-spot laser transepithelial phototherapeutic keratectomy with optical coherence tomography.

J Cataract Refract Surg

From the Center for Ophthalmic Optics and Lasers (Li, Yokogawa, Tang, Chamberlain, Zhang, Huang), Casey Eye Institute and Department of Ophthalmology, Oregon Health and Science University, Portland, Oregon, USA; and the Department of Ophthalmology (Yokogawa), Kanazawa University Graduate School of Medical Science, Kanazawa, Japan. Electronic address:

Published: April 2017

Purpose: To analyze transepithelial phototherapeutic keratectomy (PTK) results using optical coherence tomography (OCT) and develop a model to guide the laser dioptric and depth settings.

Setting: Casey Eye Institute, Portland, Oregon, USA.

Design: Prospective nonrandomized case series.

Methods: Patients with superficial corneal opacities and irregularities had transepithelial PTK with a flying-spot excimer laser by combining wide-zone myopic and hyperopic astigmatic ablations. Optical coherence tomography was used to calculate corneal epithelial lenticular masking effects, guide refractive laser settings, and measure opacity removal. The laser ablation efficiency and the refractive outcome were investigated using multivariate linear regression models.

Results: Twenty-six eyes of 20 patients received PTK to remove opacities and irregular astigmatism due to scar, dystrophy, radial keratotomy, or previous corneal surgeries. The uncorrected distance visual acuity and corrected distance visual acuity were significantly improved (P < .01) by 3.7 Snellen lines and 2.0 Snellen lines, respectively, to a mean of 20/41.2 and 20/22.0, respectively. Achieved laser ablation depths were 31.3% (myopic ablation) and 63.0% (hyperopic ablation) deeper than the manufacturer's nomogram. The spherical equivalent of the corneal epithelial lenticular masking effect was 0.73 diopter ± 0.61 (SD). The refractive outcome highly correlated to the laser settings and epithelial lenticular masking effect (Pearson R = 0.96, P < .01). The ablation rate of granular dystrophy opacities appeared to be slower. Smoothing ablation under masking fluid was needed to prevent focal steep islands in these cases.

Conclusions: The OCT-measured ablation depth efficiency could guide opacity removal. The corneal epithelial lenticular masking effect could refine the spherical refractive nomogram to achieve a better refractive outcome after transepithelial ablation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5443118PMC
http://dx.doi.org/10.1016/j.jcrs.2017.03.004DOI Listing

Publication Analysis

Top Keywords

optical coherence
12
coherence tomography
12
transepithelial phototherapeutic
8
phototherapeutic keratectomy
8
distance visual
8
visual acuity
8
laser
5
guiding flying-spot
4
flying-spot laser
4
laser transepithelial
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!