Wash-in and wash-out of sevoflurane in a test-lung model: A comparison between Aisys and FLOW-i.

F1000Res

Department of Anaesthesia & Intensive Care, Institution for Clinical Science, Karolinska Institutet, Danderyds University Hospital, Stockholm, 182 88 , Sweden.

Published: March 2017

Modern anaesthesia workstations are reassuringly tight and are equipped with effective gas monitoring, thus providing good opportunities for low/minimal flow anaesthesia. A prerequisite for effective low flow anaesthesia is the possibility to rapidly increase and decrease gas concentrations in the circle system, thereby controlling the depth of anaesthesia.  We studied the wash-in and wash-out of sevoflurane in the circle system with fixed fresh gas flow and vaporizer setting. We compared two modern anaesthesia work stations, the Aisys (GE, Madison, WI, USA) and FLOW-i (Maquet, Solna, Sweden) in a test lung model.  : We found fresh-gas flow to have, as expected, a major influence on wash-in, as well as wash-out of sevoflurane. The wash-in time to reach a stable circle 1 MAC (2.1%) decreased from an average of 547 ± 83 seconds with a constant fresh gas flow of 300 ml/min and vaporizer setting of 8%, to a mean of 38 ± 6 seconds at a fresh gas flow of 4 L/min. There were only minor differences between the two works-stations tested; the Aisys was slightly faster at both 300 and 4 L/min flow. Time to further increase circle end-tidal concentration from 1-1.5 MAC showed likewise significant associations to fresh gas and decreased from 330 ± 24 seconds at 300 ml/min. to less than a minute at constant 4 L/min (17 ± 11 seconds), without anaesthetic machine difference. Wash-out was also fresh gas flow dependent and plateaued at 7.5 L/min.  : Circle system wash-in and wash-out show clear fresh gas dependency and varies somewhat between the Aisys and Flow-i. The circle saturation, reaching 1 MAC end-tidal or increasing from 1-1.5 MAC can be achieved with both work-stations within 1.5 minutes at a constant fresh gas flow of 2 and 4 L/min. Wash-out plateaued at 7.5 L/min.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5414818PMC
http://dx.doi.org/10.12688/f1000research.11255.2DOI Listing

Publication Analysis

Top Keywords

fresh gas
28
gas flow
20
wash-in wash-out
12
wash-out sevoflurane
12
circle system
12
gas
9
flow
9
aisys flow-i
8
modern anaesthesia
8
flow anaesthesia
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!