Objectives: To compare the lung and breast dose associated with three chest protocols: standard, organ-based tube current modulation (OBTCM) and fast-speed scanning; and to estimate the error associated with organ dose when modelling the longitudinal (z-) TCM versus the 3D-TCM in Monte Carlo simulations (MC) for these three protocols.
Method: Five adult and three paediatric cadavers with different BMI were scanned. The CTDI of the OBTCM and the fast-speed protocols were matched to the patient-specific CTDI of the standard protocol. Lung and breast doses were estimated using MC with both z- and 3D-TCM simulated and compared between protocols.
Results: The fast-speed scanning protocol delivered the highest doses. A slight reduction for breast dose (up to 5.1%) was observed for two of the three female cadavers with the OBTCM in comparison to the standard. For both adult and paediatric, the implementation of the z-TCM data only for organ dose estimation resulted in 10.0% accuracy for the standard and fast-speed protocols, while relative dose differences were up to 15.3% for the OBTCM protocol.
Conclusion: At identical CTDI values, the standard protocol delivered the lowest overall doses. Only for the OBTCM protocol is the 3D-TCM needed if an accurate (<10.0%) organ dosimetry is desired.
Key Points: • The z-TCM information is sufficient for accurate dosimetry for standard protocols. • The z-TCM information is sufficient for accurate dosimetry for fast-speed scanning protocols. • For organ-based TCM schemes, the 3D-TCM information is necessary for accurate dosimetry. • At identical CTDI , the fast-speed scanning protocol delivered the highest doses. • Lung dose was higher in XCare than standard protocol at identical CTDI .
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00330-017-4863-3 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!