Optical quality of rotationally symmetrical contact lenses derived from their power profiles.

Cont Lens Anterior Eye

Department of Optics, and Optometry, and Vision Sciences, University of Valencia, Burjasot, Spain.

Published: October 2017

Purpose: To present a methodology for evaluating the optical quality of rotationally symmetrical contact lenses (CLs) from a single power profile.

Methods: Simulated rotationally symmetrical power profiles corresponding to different CLs designs (monofocal, two-zone center-near bifocal, and four-zone center-distance bifocal) were used to calculate the wavefront error profile by means of numerical integration. Then, each lens wavefront error profile was spun around the center to obtain the lens wavefront error surface. From the surface, monochromatic optical transfer functions (OTF), simulated images and the visual Strehl ratio based on the OTF (VSOTF) were obtained for different distances and pupil sizes (3 and 5.5mm) after performing a through-focus.

Results: VSOTF variations, taking into account both vergence and pupil size, were presented for the three CLs designs. The monofocal design showed excellent optical quality only for far vision, whereas the bifocal designs exhibited good optical quality for far and near vision. Modulation transfer function (MTF) from each lens design, pupil size, and work distance agreed with the previous results.

Conclusions: The methodology presented here allows for a rapid and thorough assessment of the optical quality of rotationally symmetrical CLs by means of optical quality metrics, with a special interest in simultaneous image contact lenses. This methodology may be useful for choosing the most suitable lens for each subject's visual demands.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clae.2017.05.005DOI Listing

Publication Analysis

Top Keywords

optical quality
24
rotationally symmetrical
16
quality rotationally
12
contact lenses
12
wavefront error
12
symmetrical contact
8
power profiles
8
cls designs
8
designs monofocal
8
error profile
8

Similar Publications

Simulations of the Potential for Diffraction Enhanced Imaging at 8 keV using Polycapillary Optics.

Biomed Phys Eng Express

January 2025

Physics Department, University at Albany, 1400 Washington Ave, Albany, New York, 12222-0100, UNITED STATES.

Conventional x-ray radiography relies on attenuation differences in the object, which often results in poor contrast in soft tissues. X-ray phase imaging has the potential to produce higher contrast but can be difficult to utilize. Instead of grating-based techniques, analyzer-based imaging, also known as diffraction enhanced imaging (DEI), uses a monochromator crystal with an analyzer crystal after the object.

View Article and Find Full Text PDF

End-range movements are among the most demanding but least understood in the sport of tennis. Using male Hawk-Eye data from match-play during the 2021-2023 Australian Open tournaments, we evaluated the speed, deceleration, acceleration, and shot quality characteristics of these types of movement in men's Grand Slam tennis. Lateral end-range movements that incorporated a change of direction (CoD) were identified for analysis using k-means (end-range) and random forest (CoD) machine learning models.

View Article and Find Full Text PDF

Purpose: To evaluate the efficacy and safety of intense pulsed light (IPL) combined with meibomian gland expression (MGX) for the treatment of dry eye disease and meibomian gland dysfunction associated with chronic Stevens-Johnson syndrome and toxic epidermal necrolysis.

Methods: This prospective noncomparative interventional study included 29 patients (58 eyes) who underwent 3 sessions of IPL and MGX at 2-week intervals. Subjective symptoms (ocular surface disease index score) and objective dry eye tests: matrix metalloproteinase 9, tear meniscus height, bulbar redness score, tear film lipid layer thickness (LLT), Schirmer I test, conjunctival and corneal staining, meibomian gland loss, MGX score [meibomian gland score (MGS)], and tear break-up time were assessed at the baseline and after 4, 8, and 12 weeks.

View Article and Find Full Text PDF

Purpose: To report on optical coherence tomography angiography (OCTA) in patients with a type 1 Boston keratoprosthesis (KPro) and determine its feasibility through assessment of imaging artifacts.

Methods: KPro and non-KPro subjects were matched for age, gender, and glaucoma diagnosis. OCTA images of the peripapillary optic nerve were obtained, reviewed by 2 readers masked to the diagnosis for artifacts and usability, and used for microvascular measurements.

View Article and Find Full Text PDF

COVID-19 disease, triggered by SARS-CoV-2 virus infection, has led to more than 7.0 million deaths worldwide, with a significant fraction of recovered infected people reporting postviral symptoms. Smart surfaces functionalized with nanoparticles are a powerful tool to inactivate the virus and prevent the further spreading of the disease.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!