Purpose: Exit 2D detectors are widely used in clinics as a tool for pre- treatment field verification. It is desired to have accurate modeling of the detector dose for each IMRT plan with patient geometry for in-vivo delivery verification. We propose a novel hybrid of model and measurement based methods to estimate the detector dose using the information from TPS and plan/verification CT.

Methods: Our approach is based on the generalized equivalent field size (GEFS) method. It requires two commissioning tables for various square fields (l×l, 2×2, …40×40): the percent depth dose (PDD) table and the detector correction factor (DDCF) table. PDDs are retrieved from the treatment planning system (TPS), and DDCFs are reconstructed from measurement with various field sizes and air gaps (from 5 cm to 50 cm). GEFS models the detector point dose as the superposition of annular contribution of the fluence map, which is retrieved from the TPS. Correction on the radiological path length is calculated through ray-tracing the patient CT. Corrections on the air gap between the couch and detector and detector response are applied via table lookup on PDD and DDCF.

Results: We validated the proposed method using TPS with extended geometry and direct clinic measurements for both regular and IMRT fields, various phantom and patient geometry. For all calculations, more than 98% of pixels pass the gamma index with criteria of 3%, 3mm. Each calculation took only a few seconds on a single PC.

Conclusions: We proposed a novel detector dose calculation method that can be applied for arbitrary IMRT field and arbitrary patient geometry. The calculation is simple and fast and when compared with detector measurement during IMRT treatment, makes in- vivo delivery verification and dose reconstruction feasible.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4736147DOI Listing

Publication Analysis

Top Keywords

detector dose
16
delivery verification
12
patient geometry
12
detector
9
novel detector
8
dose calculation
8
vivo delivery
8
dose
7
imrt
5
we-e-brb-05 novel
4

Similar Publications

Rhanterium Epapposum Essential Oil and Its Primary Compounds Control Infection, Inflammation, and Serum Electrolyte Imbalance in Mice with Giardiasis.

Acta Parasitol

January 2025

Department of Medical Laboratories, College of Applied Medical Sciences, Shaqra University, Ad Dawadimi, Shaqra, 17464, Saudi Arabia.

Purpose: The present experimental study seeks to evaluate the in vitro and in vivo effects, as well as the potential mechanisms of action, of Rhanterium epapposum essential oil (REE) and its main constituents against Giardia lamblia infection.

Methods: The analysis of REE was performed using the Gas Chromatography-Mass Spectrometry (GC-MS) detector. The in vitro effects of REE and its main constituents on viability of G.

View Article and Find Full Text PDF

This study assesses the activity concentrations of the radionuclides 238U, 232Th, and 40K in soil samples collected from Wolaita Sodo town, located in the Southern Nations, Nationalities, and Peoples' (SNNP) Region, Ethiopia. A gamma-ray spectrometer equipped with a NaI(Tl) detector was used for the measurements. The concentrations of 238U, 232Th, and 40K varied from 3.

View Article and Find Full Text PDF

Objective: The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).

Methods: The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed, including comparison of settings for exposure control software, different magnification, collimation configurations and detector distance.

View Article and Find Full Text PDF

The increasing use of recreational nitrous oxide ([Formula: see text]O) in the Netherlands and its link to traffic accidents highlights the need for reliable detection methods for law enforcement. This study focused on ex vivo detection of [Formula: see text]O in exhaled breath and examining its persistence in the human body. Firstly, a low-cost portable infrared based detector was selected and validated to detect [Formula: see text]O in air.

View Article and Find Full Text PDF

Purpose: To evaluate and compare lung nodules' image quality and radiation dose exposure using photon-counting computed tomography (PC-CT) and conventional energy-integrating detector computed tomography (EID-CT).

Methods: Protocol pre-registration was performed a priori at (https://osf.io/krj5y/).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!