SU-E-J-06: A Time Dependence Analysis of CBCT Image Quality and Mechanical Stability.

Med Phys

South Carolina Oncology Associates, Columbia, SC.

Published: June 2012

Purpose: To quantify the change, if any, in flexmap correction factors and image quality with the XVI system over a course of several years and from these results, assess their clinical impact.

Methods: Flexmap, a calibration procedure which corrects for imperfect gantry rotation for cone-beam CT reconstruction, and image quality tests were performed on three Elekta Synergy linacs equipped with XVI. Data was collected per month over three years. U and V values, corresponding to lateral and longitudinal shifts respectively, were acquired through the XVI software. Image quality parameters were obtained through CT imaging of the Catphan 500®. For each reconstruction, pixel values for low density polyethylene (LDPE) and polystyrene materials were recorded.

Results: For all three linacs, analysis of the flexmap showed a significant change in the U factor for both month-to-month comparisons and comparisons between machines. The V correction factor exhibited a small variation month to month, and showed a slight, gradual increase over time (0.2 +/-0.08 mm). Image quality analysis showed a near consistent decrease (5-10%) in LDPE and polystyrene. Despite this decrease in pixel values, the ratio of the two pixel values remained constant, thus a similar decreasing trend in contrast was not observed.

Conclusions: Analysis of monthly flexmap calibration showed the general monthly change in correction shifts and their general trend over several years. For image quality, our research exhibited roughly 0.5% per month decrease in pixel values of the Catphan®. Our results imply that CBCT images obtained from XVI are not appropriate for treatment planning and despite the decrease in panel response over time, image quality with respect to contrast will remain within acceptable clinical standards. Future studies may be carried out to assess any correlation between image quality and XVI source strength.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4734838DOI Listing

Publication Analysis

Top Keywords

image quality
32
pixel values
16
image
8
quality
8
quality xvi
8
flexmap calibration
8
ldpe polystyrene
8
despite decrease
8
decrease pixel
8
xvi
5

Similar Publications

Ultrasound blood flow imaging plays a crucial role in the diagnosis of cardiovascular and cerebrovascular diseases. Conventional ultrafast ultrasound plane-wave imaging techniques have limited capabilities in microvascular imaging. To enhance the quality of blood flow imaging, this study proposes a microbubble-based H-Scan ultrasound imaging technique.

View Article and Find Full Text PDF

Assessing water quality restoration measures in Lake Pampulha (Brazil) through remote sensing imagery.

Environ Sci Pollut Res Int

January 2025

LEESU, Ecole des Ponts Paris Tech, UPEC, AgroParisTech, F-77455 Marne-la-Vallée, Paris, France.

Urban reservoirs are frequently exposed to impacts from high population density, polluting activities, and the absence of environmental control measures and monitoring. In this study, we investigated the use of satellite imagery to assess restoration measures and support decision-making in a hypereutrophic urban reservoir. Since 2016, Lake Pampulha (Brazil) has undergone restoration measures, including the application of Phoslock®, to mitigate its poor water quality conditions.

View Article and Find Full Text PDF

Pathology has benefited from the rapid progress of image-digitizing technology during the last decade. However, the application of digital whole slide images (WSI) in forensic pathology still needs to be improved. WSI validation is crucial to ensure diagnostic performance, at least equivalent to glass slides and light microscopy.

View Article and Find Full Text PDF

Objectives: To determine the value of preoperative magnetic resonance imaging (MRI) in predicting macrotrabecular-massive hepatocellular carcinoma (MTM-HCC).

Materials And Methods: A search was conducted on PubMed, Web of Science, Cochrane Library databases, and Embase for studies evaluating the performance of MRI in assessing MTM-HCC. The quality assessment of diagnostic studies (QUADAS-2) tool was used to assess the risk of bias.

View Article and Find Full Text PDF

Cardiovascular imaging in children with cardiac implantable electronic devices.

Pediatr Radiol

January 2025

Department of Cardiology, Boston Children's Hospital and Department of Pediatrics, Harvard Medical School, Boston, MA, 02115, USA.

The number of children with cardiac implantable electronic devices (CIEDs) is increasing at a time of rapid growth in cardiac magnetic resonance (MR) and cardiac computed tomography (CT) utilization. The presence of CIEDs poses challenges with respect to imaging safety and quality. A thoughtful approach to cardiovascular imaging in patients with CIEDs begins with an awareness of the clinical indications to determine the most appropriate imaging modality.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!