Diagnosis of preeclampsia and fetal growth restriction with the sFlt-1/PlGF ratio: Diagnostic accuracy of the automated immunoassay Kryptor®.

Pregnancy Hypertens

Department of Laboratory Medicine, Clinical Chemistry, and Pathobiochemistry, Charité Universitätsmedizin Berlin, Germany; Labor Berlin - Charité Vivantes GmbH, Berlin, Germany.

Published: April 2017

Objective: We aimed to characterize the diagnostic accuracy of the Kryptor® assay for sFlt-1 and PlGF in maternal serum samples of uneventful singleton pregnancies and subjects with preeclampsia (PE) and PE-related outcomes such as fetal growth restriction (FGR). Longitudinal reference ranges of the sFlt-1 and PlGF level in the course of normal pregnancies were generated.

Methods: A cohort of subjects with PE and PE-related outcomes including FGR in the third trimester was compared to a cohort of women with uneventful outcome. Serum levels of sFlt-1, PlGF level as well as the sFlt-1/PlGF ratio was analysed with the Kryptor® assay and compared between the case- and control groups. Cut-off values were generated and diagnostic accuracy examined.

Results: Longitudinal reference ranges of the sFlt-1 and PlGF level in healthy pregnancies were in line with those levels measured with other immunoassays. Comparison of the sFlt-1/PlGF ratio between PE-related outcomes including FGR or PE and healthy controls showed a high diagnostic accuracy with an area under the curve (AUC) of 0.917 for PE-related outcomes and 0.919 for PE.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.preghy.2017.02.005DOI Listing

Publication Analysis

Top Keywords

diagnostic accuracy
16
sflt-1 plgf
16
pe-related outcomes
16
sflt-1/plgf ratio
12
plgf level
12
fetal growth
8
growth restriction
8
kryptor® assay
8
longitudinal reference
8
reference ranges
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!