Aim: The performance of glucagon and GLP-1 immunoassays is often poor, but few sensitive LC-MS/MS methods exist as alternatives.
Experimental: A multiplexed LC-MS/MS method using a 2D extraction technique was developed.
Results: The method was established for the quantitation of endogenous glucagon (LLOQ: 15 pg/ml) and dosed GLP-1 (LLOQ: 25 pg/ml) in human plasma, and is the first such method avoiding immunoenrichment. Specificity of endogenous glucagon quantitation was assured using a novel approach with a supercharging mobile phase additive to access a sensitive qualifier SRM. Endogenous glucagon concentrations were within the expected range, and showed good reproducibility after extended sample storage. A cross-validation against established immunoassays using physiological study samples demonstrated some similarities between methods.
Conclusion: The LC-MS/MS method offers a viable alternative to immunoassays for quantitation of endogenous glucagon, dosed glucagon and/or dosed GLP-1.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4155/bio-2017-0021 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!