Objectives: The aim of the study was to assess the influence of cone beam computed tomography (CBCT) acquisition parameters on the evaluation of mandibular third molars and their relationship to the mandibular canal.

Study Design: Eight dry human mandibles with 13 mandibular third molars were scanned with one CBCT unit. Voxel size (0.2 and 0.3 mm), field of view (FOV) size (12 × 8.5 cm and 5 × 5 cm), and number of basis images (450 and 720) were the variables studied. Two examiners evaluated the images, and the resulting data were compared through McNemar, McNemar-Bowker, and Student t tests. Additionally, dosimetry was determined for all protocols tested, and radiation doses were compared through analysis of variance.

Results: The variables did not influence evaluation of mandibular third molars, except for voxel size, when assessing contact between the tooth and the mandibular canal (P = .021). Although FOV and number of basis images affected radiation dose, voxel size did not.

Conclusions: FOV size and number of basis images did not influence the evaluation of mandibular third molars and their relationship to the mandibular canal in the CBCT unit used. Conversely, smaller voxel size affected the assessment of contact between the tooth and the mandibular canal. In units in which voxel size does not influence radiation dose, the most appropriate CBCT protocol is the one using a smaller voxel size and delivering the lowest radiation dose to the patient.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.oooo.2017.03.008DOI Listing

Publication Analysis

Top Keywords

voxel size
24
mandibular third
20
third molars
20
evaluation mandibular
16
number basis
12
basis images
12
mandibular canal
12
radiation dose
12
mandibular
9
acquisition parameters
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!