Inadequacy of current pediatric epinephrine autoinjector needle length for use in infants and toddlers.

Ann Allergy Asthma Immunol

Department of Emergency Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts.

Published: June 2017

Background: Epinephrine injection represents the standard of care for anaphylaxis treatment. It is most effective if delivered intramuscularly, whereas inadvertent intraosseous injection may be harmful. The needle length in current pediatric epinephrine autoinjectors (EAIs) is 12.7 mm; however, the ideal needle length for infants and toddlers weighing less than 15 kg is unknown.

Objective: To determine the skin-to-bone distance (STBD) and skin-to-muscle distance (STMD) at baseline and after simulated EAI application in infants and toddlers (weighing 7.5-15 kg).

Methods: Study participants recruited from 2 North American allergy clinics underwent baseline and compression (10-lb pressure) ultrasonography of the anterolateral thigh with a modified ultrasound transducer mimicking the footprint and maximum pressure application of an EAI device. Ultrasound images, with clinical data masked, were analyzed offline for STBD and STMD in short-axis approach.

Results: Of 53 infants (mean age, 18.9 months; 54.7% male; 81.1% white; mean weight, 11.0 kg), 51 had adequate images for short-axis STBD measurements. In these infants, the mean (SD) baseline STBD was 22.4 (3.8 mm), and the mean (SD) STMD was 7.9 (1.7) mm. With 10-lb compression, the mean (SD) STBD was 13.3 (2.1) mm, and the mean (SD) STMD was 6.3 (1.2) mm. An EAI with a needle length of 12.7 mm applying 10-lb pressure could strike the bone in 43.1% of infants and toddlers in this cohort.

Conclusion: Our data suggest that the optimal EAI needle length for infants and toddlers weighing 7.5 to 15 kg should be shorter than the needle length in currently available pediatric EAIs to avoid accidental intraosseous injections.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.anai.2017.03.017DOI Listing

Publication Analysis

Top Keywords

needle length
24
infants toddlers
20
length infants
12
toddlers weighing
12
current pediatric
8
pediatric epinephrine
8
10-lb pressure
8
eai needle
8
infants
7
needle
6

Similar Publications

Background And Objective: During the last 200 years, there have been many changes in the way of performing endodontic treatment. The increased demand from patients for saving their teeth has led to the development of various innovative equipment and advances in the field of material sciences. Thus, the standard protocol of endodontic treatment has undergone several modifications.

View Article and Find Full Text PDF

What Is the Optimal Geometry of Dissolving Microneedle Arrays? A Literature Review.

Pharmaceutics

January 2025

Department of Pharmaceutical Technology and Biopharmacy, University of Groningen, 9713 AV Groningen, The Netherlands.

The application of dissolving microneedle arrays (DMNAs) is an emerging trend in drug and vaccine delivery as an alternative for hypodermic needles or other less convenient drug administration methods. The major benefits include, amongst others, that no trained healthcare personnel is required and that the recipient experiences hardly any pain during administration. However, for a successful drug or vaccine delivery from the DMNA, the microneedles should be inserted intact into the skin.

View Article and Find Full Text PDF

Microneedle(MN)-based drug delivery is one of the potential approaches to overcome the limitations of oral and hypodermic needle delivery. An in silico model has been developed for hollow microneedle (HMN)-based drug delivery in the skin and its subsequent absorption in the blood and tissue compartments in the presence of interstitial flow. The drug's reversible specific saturable binding to its receptors and the kinetics of reversible absorption across the blood and tissue compartments have been taken into account.

View Article and Find Full Text PDF

Laparo-endoscopic hernia surgery is recommended by various international bodies. However, its uptake by general surgeon is low. We aim to assess the impact of Three Dimensional (3D) endovision system in learning laparoscopic transabdominal preperitoneal (TAPP) repair of groin hernia and transferability of skills acquired from 3D to the Two Dimensional (2D) environment.

View Article and Find Full Text PDF

Design of a wasp-inspired biopsy needle capable of self-propulsion and friction-based tissue transport.

Front Bioeng Biotechnol

January 2025

Department of BioMechanical Engineering, Bio-Inspired Technology Group, Faculty of Mechanical Engineering, Delft University of Technology, Delft, Netherlands.

Percutaneous pancreatic core biopsy is conclusive but challenging due to large-diameter needles, while smaller-diameter needles used in aspiration methods suffer from buckling and clogging. Inspired by the ovipositor of parasitic wasps, which resists buckling through self-propulsion and prevents clogging via friction-based transport, research has led to the integration of these functionalities into multi-segment needle designs or tissue transport system designs. This study aimed to combine these wasp-inspired functionalities into a single biopsy needle by changing the interconnection of the needle segments.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!