Rationale: The zero-profile, standalone device (Zero-P, Synthes GmbH, Switzerland) has been reported to be an effective and safe treatment method with similar clinical outcomes compared with plate. Instrumental complications concerning Zero-P have been little reported. Considering the rarity, we present this amazing case to share with our spinal surgeons and instrument specialists.
Patient Concerns: A 46-year-old man patient presented to our hospital with neck and shoulders pain for 23 years, numbness and weak-ness of right hand for 6 months.
Diagnoses: Hypoesthesia in the right C6 and C7 roots distribution, myodynamia weakness of the right little finger was detected from physical examination. Two-level anterior cervical decompression and fusion (ACDF) using the Zero-P was performed via a classic right Smith-Robinson approach after induction of general anesthesia. Three months postoperative x-rays showed a good position of the implant. Six months postoperative x-rays showed a locking screw at the segment C6/7 pulled out. The patient was diagnosed as screw pullout after ACDF.
Interventions: The patient was treated conservatively with regular follow-up as he was asymptomatic and no evidence of esophageal perforation was detected.
Outcomes: The patient was followed again and 24 months postoperative x-rays also showed the pulled-out locking screw had re-screwed spontaneously. The patient was noticed that a revision surgery was needed if symptoms occur. At present bony union is not reached but he is still asymptomatic.
Lessons: Pulled-out screws re-screwed spontaneously are rare but it does occur. Insertion angle may affect the stability of the Zero-P device, and the repeated micro-motion may be the critical reason of the screw pull-out and re-screwing. The management of screws pull-out after ACDF remains individualized and a revision surgery is not necessary for every patient. Conservative treatment such as orthosis and regular follow-ups may be suitable for some asymptomatic patients.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5419933 | PMC |
http://dx.doi.org/10.1097/MD.0000000000006827 | DOI Listing |
Eur J Orthop Surg Traumatol
December 2024
Department of Orthopedics and Trauma, Universidade Federal Fluminense (UFF), Niterói, Rio de Janeiro, Brazil.
Purpose: Although several techniques have been described for bent intramedullary nail removal, there is no universally accepted strategy. We hypothesized that a device based on the action principle of a three-point bend fixture could facilitate extraction of bent intramedullary nails; this paper describes its design and experimental testing.
Methods: Five large synthetic left femurs and five steel intramedullary nails were used.
BMC Musculoskelet Disord
December 2024
Department of Anatomy, Guangdong Provincial Key Laboratory of Digital Medicine and Biomechanics, Guangdong Engineering Research Center for Translation of Medical 3D Printing Application, National Virtual & Reality Experimental Education Center for Medical Morphology, School of Basic Medical Sciences, Southern Medical University, No.1023, South Shatai Road, Baiyun District, Guangzhou, Guangdong, 510515, China.
Background: This study investigated the impact of higher interfragmentary compression force (IFCF) on the stability of locking plate fixation in lateral tibial plateau fractures.
Methods: Biomechanical experiments and finite element analysis (FEA) were employed to compare the performance of the AO cancellous lag screw (AOCLS) and a newly developed combined cancellous lag screw (CCLS).
Results: The results demonstrated that the CCLS provided a higher IFCF without the risk of over-screwing, significantly improving fixation stability.
World J Surg
December 2024
Department of Cardiothoracic Surgery, The Third Hospital of Shijiazhuang City, Shijiazhuang, China.
Background: Costal cartilage injuries are unappreciated, and there is a paucity of reports on fixation methods. This study aims to evaluate the safety of titanium plate internal fixation for costal cartilage injuries.
Methods: A retrospective analysis was conducted on 30 patients with costal cartilage injuries who underwent titanium plate internal fixation between April 2016 and November 2022 at our hospital.
Background: Achieving bony union in scaphoid nonunion fractures is challenging. Various bone grafts have been studied using headless compression screws (HCS) fixation. However, the impact of bone graft choice on bone healing with volar locking plates (VLP) use is less clear.
View Article and Find Full Text PDFInjury
December 2024
Department of Orthopaedics, Larnaca General Hospital, State Health Services Organisation, Larnaca, Cyprus.
The purpose of this study was to establish typical dose values at orthopaedic operating rooms of the Larnaca General Hospital (LGH). Kerma area product (KAP), fluoroscopy time (FT) and cumulative air-kerma (K) measurements were collected for 821 patients who underwent common and reproducible trauma surgery over a five-year period, with three mobile C-arm systems; two equipped with an image-intensifier and one with a flat-panel detector. Dose indices were automatically extracted from radiation dose structured reports or DICOM meta-data files archived in the PACS, using custom-made software.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!