Complications after implantation of a new-generation insertable cardiac monitor: Results from the LOOP study.

Int J Cardiol

Department of Cardiology, Rigshospitalet, Copenhagen University Hospital, Blegdamsvej 9, 2100 Copenhagen, Denmark; The Danish National Research Foundation Centre for Cardiac Arrhythmia, Blegdamsvej 3, 2200 Copenhagen, Denmark; Department of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, Blegdamsvej 3B, 2200 Copenhagen, Denmark.

Published: August 2017

Background: Insertable cardiac monitors (ICM) are leadless devices utilized in long-term monitoring of the heart rhythm. The implantation procedure of the new-generation ICMs is minimally invasive, but little experience exists regarding complications. We thus aimed to investigate adverse events (AE) according to procedure-related characteristics after implantation of a large number of new-generation ICMs.

Methods: The study population consisted of participants randomized to receive ICM in a multi-center trial. The Reveal LINQ™ ICM was implanted using provided insertion tools, either in an electrophysiology laboratory or outpatient procedure room. If device sensing was insufficient in the first subcutaneous position, one or more repositions were performed. Patients were urged to make contact if they suspected any AE. Furthermore, follow-up for safety endpoints consisted of evaluation of medical records and planned study visits.

Results: 1420 patients received an ICM, 753 (53%) in a procedure room and 667 (47%) in an electrophysiology laboratory. During a median follow-up of 499days, 9 (0.63%) and 15 (1.13%) patients experienced AEs with and without need for device explantation, respectively. In the 38 patients requiring device repositioning, more AEs requiring explantation were seen (3 (7.9%) vs. 6 (0.4%), p=0.001). Patients undergoing implantation in a procedure room had more infections (12 (1.6%) vs. 1 (0.1%), p=0.004), though no significant difference was reached in AEs requiring explantation (7 (0.9%) vs. 2 (0.3%), p=0.19).

Conclusion: The Reveal LINQ™ ICM can be inserted with a very low risk of complications, both in the traditional electrophysiology laboratory setting and in an outpatient procedure room.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijcard.2017.03.144DOI Listing

Publication Analysis

Top Keywords

procedure room
16
electrophysiology laboratory
12
insertable cardiac
8
implantation procedure
8
reveal linq™
8
linq™ icm
8
outpatient procedure
8
aes requiring
8
requiring explantation
8
icm
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!