Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer.

Int Ophthalmol

Research Department, Asociación para Evitar la Ceguera en México I.A.P, Vicente García Torres 46, Barrio San Lucas, Coyoacan, 04030, Mexico City, Mexico.

Published: June 2018

Purpose: To compare the biometric measurements obtained from the Verion Image-Guided System to those obtained by auto-refracto-keratometer in normal eyes.

Methods: This is a prospective, observational, comparative study conducted at the Asociación para Evitar la Ceguera en México I.A.P., Mexico. Three sets of keratometry measurements were obtained using the image-guided system to assess the coefficient of variation, the within-subject standard deviation and intraclass correlation coefficient (ICC). A paired Student t test was used to assess statistical significance between the Verion and the auto-refracto-keratometer. A Pearson's correlation coefficient (r) was obtained for all measurements, and the level of agreement was verified using Bland-Altman plots.

Results: The right eyes of 73 patients were evaluated by each platform. The Verion coefficient of variation was 0.3% for the flat and steep keratometry, with the ICC being greater than 0.9 for all parameters measured. Paired t test showed statistically significant differences between groups (P = 0.0001). A good correlation was evidenced for keratometry values between platforms (r = 0.903, P = 0.0001 for K1, and r = 0.890, P = 0.0001). Bland-Altman plots showed a wide data spread for all variables.

Conclusion: The image-guided system provided highly repeatable corneal power and keratometry measurements. However, significant differences were evidenced between the two platforms, and although values were highly correlated, they showed a wide data spread for all analysed variables; therefore, their interchangeable use for biometry assessment is not advisable.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10792-017-0541-3DOI Listing

Publication Analysis

Top Keywords

image-guided system
16
biometric measurements
8
measurements verion
8
verion image-guided
8
keratometry measurements
8
coefficient variation
8
correlation coefficient
8
wide data
8
data spread
8
measurements
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!