Introduction: EBUS is a well established minimally invasive diagnostic tool for mediastinal and hilar lymphadenopathy. The novel ViziShot Flex 19G needle (Olympus Respiratory America, Redmond, WA, USA) was introduced in 2015 in order to improve loaded scope flexion and to obtain larger tissue samples for analysis. The aims of this study were to assess diagnostic yield of Flex 19G needles and to present endoscopist's feedback about the endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA).

Material And Methods: The Flex 19G needles were used in patients with hilar and/or mediastinal adenopathy in two Polish pulmonology centers. Cytology smears and cell blocks (CB) were prepared. The prospective analysis was performed due to collected data.

Results: Twenty two selected patients with confirmed adenopathy on chest-CT (mean age 58 ± 12) underwent EBUS-TBNA with use of Flex 19G needles. All procedures occurred to be diagnostic for smears (yield 100%). The malignancy was found in 15 cases (68.2%), and benign adenopathy in 7 (31.8%). In 12 of 14 cases of lung cancer (yield 85.7%) CB were diagnostic for immunohistochemical and molecular staining. After puncturing nodes, especially in hilar position not extensive bleeding was observed. Comparing to standard 21/22G EBUS-TBNA endoscopists underlined better flexion of loaded scope and sample adequacy and found non-significant differences in another biopsy details.

Conclusions: The first Polish experience with use of Flex 19G EBUS-TBNA needle occurs to be similar in performance with standard technique with use of 22/21G needles and presents high diagnostic yield for lung cancer diagnostics, especially when preparing CB. A safety profile of the biopsy is acceptable.

Download full-text PDF

Source
http://dx.doi.org/10.5603/ARM.2017.0012DOI Listing

Publication Analysis

Top Keywords

flex 19g
20
19g needles
12
polish experience
8
endobronchial ultrasound-guided
8
ultrasound-guided transbronchial
8
transbronchial needle
8
needle aspiration
8
loaded scope
8
diagnostic yield
8
lung cancer
8

Similar Publications

Evolution of transbronchial needle aspiration needles: Over the last half century.

Chin Med J Pulm Crit Care Med

September 2024

Department of Pulmonary and Critical Care Medicine, Johns Hopkins University School of Medicine, Baltimore, MD 21224, USA.

Article Synopsis
  • * The introduction of endobronchial ultrasound-guided TBNA (EBUS-TBNA) has improved real-time imaging and safety of the procedure with advanced needle designs like ViziShot2 and Flex 19G.
  • * The review covers the evolution of TBNA, including successes and challenges, to help both experienced pulmonologists and new physicians improve their understanding and technique in modern TBNA practices.
View Article and Find Full Text PDF

Background: Endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is a well-established diagnostic procedure for evaluating hilar and mediastinal lymphadenopathies and is the gold standard for lung cancer diagnosis and staging. Recent studies assessed the effectiveness of the 19-G flex needle in obtaining larger EBUS-TBNA samples, and prospective small series gave similar results in terms of diagnostic yield when testing different gauge needles. The lack of homogeneity between series and the small sample size of some prospective cohorts poses a limit to the validity of those results.

View Article and Find Full Text PDF

Background: Pancreatic cystic lesions (PCLs) are increasingly found on cross-sectional imaging, with the majority having a low risk for malignancy. The added value of fine-needle aspiration (FNA) in risk stratification remains unclear. We evaluated the impact of three FNA needles on diagnostic accuracy, clinical management, and the ability to accrue fluid for tumor markers.

View Article and Find Full Text PDF

Background: Previous studies have shown that needle gauge size has no significant impact on diagnostic yield during endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA). Our objective was to determine whether cell blocks obtained via the new Flex 19G EBUS-TBNA needle would contain more cellular material based on cell area compared with those obtained from a 21G needle.

Methods: A prospective analysis of patients undergoing EBUS-TBNA at our institutions was performed.

View Article and Find Full Text PDF

Objective Endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is a minimally invasive procedure for pulmonary and mediastinal diseases. This study compared the performance of a novel 19-G needle with a 21- or 22-G needle. Patients and Methods Eleven patients at Okayama Medical Center were enrolled retrospectively between April and December 2017.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!