Objective: To investigate the relationship between the morphological patency of the eustachian tube (ET) and sound transmission via the ET.
Study Design: Retrospective chart review and model experiment.
Setting: Tertiary referral center.
Subjects: A total of 56 ears of 28 patients (9 men and 19 women, aged from 12 to 82 yr, mean 40.4 ± 21.5 yr) who underwent sonotubometric measurement using postural change as well as computed tomography (CT) of the ET in the sitting position, including 26 ears with definite patulous ET, 9 ears with possible patulous ET, and 21 ears without patulous ET.
Method: Commercial equipment for sonotubometry (JK-05A; RION Co., Ltd., Kokubunji, Tokyo, Japan) was used in the following two investigations. Retrospective survey: sound patency of 7 kHz band noise via the ET was assessed by comparison of acoustic transfer function via the ET in the sitting and forward-bending positions. Sound patency via the ET was compared with morphological patency of the ET (cross-sectional area in the narrowest portion) assessed by three-dimensional CT of the ET in the sitting position. Model experiment: effect of the ET caliber on the acoustic transfer function was examined using a simple model constructed with two truncated syringes with silicone barrels and a narrow connecting tube.
Results: Sound patency assessed by sonotubometry was well correlated with the cross-sectional area at the narrowest portion of the ET in the sitting position (r = 0.786, p < 0.001). Correlation between the sound patency via the narrow tube and the caliber of the tube was also confirmed by the model experiment.
Conclusions: Sound patency assessed by sonotubometry using 7 kHz band noise could be useful to predict the morphological severity of patulous ET.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/MAO.0000000000001413 | DOI Listing |
Neurosurg Rev
January 2025
Department of Neurosurgery, Sana Kliniken Duisburg, Academic Teaching Hospital of University Duisburg-Essen, Duisburg, Germany.
Pineal gland lesions pose a significant surgical challenge due to the deep-seated nature of the pineal gland, as well as the limited field of view, and the complex vascular anatomy. The mainstay of surgical treatment, when necessary, is always histopathological clarity and gross total resection (GTR). We evaluate the surgical outcomes for pineal gland lesions, shedding light on functional outcomes, histological findings, and surgical complications.
View Article and Find Full Text PDFPhysiol Rep
February 2025
Department of Biomedical Engineering, Toyo University, Saitama, Japan.
The present study aims to examine the effect of 4 h of continuous sitting on cerebral endothelial function, which is a crucial component of cerebral blood flow regulation. We hypothesized that 4 h of sitting may impair cerebral endothelial function similarly to how it affects lower limb vasculature. Thirteen young, healthy participants were instructed to remain seated for 4 h without moving their lower limbs.
View Article and Find Full Text PDFAnn Vasc Dis
January 2025
Department of Surgery and Clinical Science, Yamaguchi University Graduate School of Medicine, Ube, Yamaguchi, Japan.
We sought to clarify the interface pressure (IP) and its variation by applying Biflex16, a calibrated bandage, to the lower leg. In Study I, 50 participants applied a bandage to the lower leg of a single subject in two ways: first, with 50% overlap, while the calibration rectangle became a square (Application 1), and then with 50% overlap without intentional stretch (Application 2) which served as a control. In Study II, another 51 participants applied the bandage to their lower leg via Application 1.
View Article and Find Full Text PDFActa Bioeng Biomech
June 2024
1Department of Rehabilitation Medicine, Southern Medical University Nanfang Hospital, Guangzhou, China.
: The purpose of this study was to quantify the impact of smartphone use while sitting on the toilet on the spinal flexion angles and the time effect. : Measurements of the spinal flexion angles in the sagittal plane were made by thirty participants while they sat on the toilet for 10 min, using a smartphone in either one, both, or neither hand. The individual's forehead, cervical, thoracic and lumbar spinal areas were each fitted with five different inertial motion sensors.
View Article and Find Full Text PDFActa Bioeng Biomech
June 2024
2Institute of Sound and Vibration Research, University of Southampton, United Kingdom.
Transmissibility is used to assess dynamic responses of the occupant-seat system, and most studies have exclusively assessed the transmissibility from the floor to the cushion or the backrest surface with the human body. In this investigation, the vertical vibration transmitted from the floor to six specific locations both on the seat surface and the frame when the seat was fixed on three positions on the track was examined utilizing an SAE J826 manikin and 12 male adults (0.25 to 20 Hz) for a duration of 120 seconds at three vibration amplitudes.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!