Purpose: This study aims to investigate how the positions of the feeding sources of the transmit radiofrequency (RF) coil, field orientation direction with respect to the patient, and patient dimensions affect the global and local electromagnetic exposure in human body models.

Methods: Three RF coil models were implemented, namely a specific two-source (S2) feed and two multisource feed configurations: generic 32-source (G32) and hybrid 16-source (H16). Thirty-two feeding conditions were studied for the S2, whereas two were studied for the G32 and H16. The study was performed using five human body models. Additionally, for two of the body models, the case of a partially implanted lead was evaluated.

Results: The results showed an overall variation due to coil feeding conditions of the whole-body specific absorption rate (SAR) of less than 20%, but deviations up to 98% of the magnitude of the electric field tangential to a possible lead path. For the analysis with the partially implanted lead, a variation of local SAR at the tip of the lead of up to 60% was observed with respect to feed position and field orientation direction.

Conclusion: The results of this study suggest that specific information about feed position and field orientation direction must be considered for an accurate evaluation of patient exposure. Magn Reson Med 79:1135-1144, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5810925PMC
http://dx.doi.org/10.1002/mrm.26703DOI Listing

Publication Analysis

Top Keywords

human body
12
body models
12
field orientation
12
global local
8
local electromagnetic
8
exposure human
8
orientation direction
8
feeding conditions
8
partially implanted
8
implanted lead
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!