Background: Many clinical laboratories use a clotting rate assay according to Clauss for the determination of fibrinogen in citrated plasma. The aim of the present study was to assess the commutability of the current International Standard for fibrinogen (coded 09/264), three commercial fibrinogen standards, and 10 freeze-dried plasma quality control samples from various sources.

Methods: Clotting rate assays according to Clauss were performed on three automated instruments (Sysmex CA1500, STA-Rack Evolution and ACL-Top 700), using three commercial thrombin reagents (Siemens, Stago, and Instrumentation Laboratory). Relationships between the results obtained with the three instruments were determined with 25 fresh-frozen plasma samples obtained from patients. The deviations of the assay results obtained with the freeze-dried samples were compared with the deviations obtained with the fresh-frozen samples, according to approved CLSI guideline C53A.

Results: Freezing and thawing had no influence on the assay results. There were significant differences in the mean assay results (fibrinogen, g/L) for the fresh-frozen plasma samples between the three automated instruments: 2.51 (STA-Rack Evolution), 2.25 (ACL-Top 700) and 2.20 (Sysmex CA1500). Similar differences were observed for several freeze-dried plasma samples. Some freeze-dried plasma samples, including the International Standard, were out of the 95% confidence interval for the relationship between STA-Rack Evolution and Sysmex CA1500.

Conclusions: Some freeze-dried plasmas including the international standard for fibrinogen are not commutable among automated instruments for fibrinogen clotting rate assays according to Clauss. Our results have consequences for all interested parties in the traceability chain (WHO, industry, external quality assessment schemes, clinical laboratories).

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2016-1088DOI Listing

Publication Analysis

Top Keywords

plasma samples
16
clotting rate
12
international standard
12
freeze-dried plasma
12
automated instruments
12
sta-rack evolution
12
quality control
8
samples
8
control samples
8
clinical laboratories
8

Similar Publications

Background: Chronic lymphocytic leukemia (CLL) is a common hematologic malignancy. Although previous research has explored associations between plasma proteins and CLL, the causal relationships remain unclear. This study used Mendelian randomization (MR) to investigate the causal relationship between 7156 plasma proteins and CLL risk.

View Article and Find Full Text PDF

Background: Numerous studies have highlighted the role of oxidative stress in Alzheimer's disease (AD) development. Yet, the alignment of systemic and central oxidative stress biomarkers is unclear across diverse populations in the AD continuum. This study aims to assess protein damage levels in plasma and cerebrospinal fluid (CSF) within the AD continuum.

View Article and Find Full Text PDF

Gliomas are the most common brain tumor type in children and adolescents. To date, diagnosis and therapy monitoring for these tumors rely on magnetic resonance imaging (MRI) and histopathological as well as molecular analyses of tumor tissue. Recently, liquid biopsies (LB) have emerged as promising tool for diagnosis and longitudinal tumor assessment potentially allowing for a more precise therapeutic management.

View Article and Find Full Text PDF

Most of advanced non-small cell lung cancer (NSCLC) patients will experience tumor progression with immunotherapy (IO). Preliminary data suggested an association between high plasma HGF levels and poor response to IO in advanced NSCLC. Our study aimed to evaluate further the role of the HGF/MET pathway in resistance to IO in advanced NSCLC.

View Article and Find Full Text PDF

Background: Apolipoprotein E (ApoE) exists in three protein isoforms: E2, E3, and E4, which differ by only one or two amino acids. These slight differences profoundly effect protein structure and function, allowing each isoform to differentially impact Alzheimer's Disease (AD) risk. Relative to the most common E3 isoform, E4 dramatically increases risk, while E2 confers a substantial decrease in risk.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!