Comparison of Gravity Flow Rates Between ENFit and Legacy Feeding Tubes.

JPEN J Parenter Enteral Nutr

Division of Endocrinology, Diabetes, Metabolism, and Nutrition, Mayo Clinic, Rochester, Minnesota, USA.

Published: March 2018

Background: Misconnections between enteral supplies and other access devices have led to significant morbidity and mortality. To reduce misconnections, a standard small-bore connector has been developed (International Organization for Standards 80369-8; ENFit). The full impact of transition to this connector is not known, however.

Method: Working with major manufacturers and Food and Drug Administration, we obtained ENFit and comparative legacy tubes of variable sizes (low-profile, 14F, 18F, 20F, and 24F balloon gastrostomies). Gravity enteral feeding was simulated with an empty bolus syringe attached to the feeding tube to be tested. The tube was clamped and filled to the 60-mL mark with liquid (water, Jevity 1 Cal, Isosource HN, Isosource 1.5 Cal, Two Cal HN, and Nourish). The clamp was released, and time for formula to leave the syringe was recorded.

Results: There was no difference in flow rate between the aggregate legacy and ENFit tubes for the low-profile, 18F, and 20F sizes. The ENFit 14F tubes had a lower flow rate vs the legacy tubes, largely due to the low flow rates seen with the 1 ENFit tube. Similarly, 24F ENFit tubes with some formulas yielded lower flow rates as opposed to legacy.

Conclusion: Overall, for the low-profile, 18F, and 20F sizes, the ENFit tubes had similar flow rates when compared with the legacy tubes. For the 14F and 24F sizes, the flow rate of ENFit tubes was significantly lower, which could result in longer EN delivery for patients who are using these tubes to provide gravity feeding.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0148607117703959DOI Listing

Publication Analysis

Top Keywords

flow rates
16
enfit tubes
16
legacy tubes
12
18f 20f
12
flow rate
12
tubes
10
enfit
9
rates enfit
8
low-profile 18f
8
20f sizes
8

Similar Publications

Bipolar disorder (BD) involves altered reward processing and decision-making, with inconsistencies across studies. Here, we integrated hierarchical Bayesian modelling with magnetoencephalography (MEG) to characterise maladaptive belief updating in this condition. First, we determined if previously reported increased learning rates in BD stem from a heightened expectation of environmental changes.

View Article and Find Full Text PDF

An online segmented continuous flow analysis system for rapid determining chemical oxygen demand in seawater to assess organic pollution levels.

Mar Pollut Bull

January 2025

Coastal Zone Ecological Environment Monitoring Technology and Equipment Shandong Engineering Research Center, CAS Key Laboratory of Coastal Zone Environmental Processes and Ecological Remediation, Yantai Institute of Coastal Zone Research, Chinese Academy of Sciences, Yantai, Shandong 264003, China; University of Chinese Academy of Sciences, Beijing 100049, China; Center for Ocean Mega-Science, Chinese Academy of Sciences, Qingdao, Shandong 266071, China; Shandong Key Laboratory of Coastal Environmental Processes, Yantai, Shandong 264003, China. Electronic address:

By integrating ultraviolet (UV) photocatalytic oxidation digestion with segmented continuous flow analysis technology, an online measurement method and analysis system for the alkaline chemical oxygen demand (COD) in seawater, based on the color-change reaction of potassium permanganate, has been established. This represents the first application of UV photocatalytic oxidation technology in the measurement of COD in seawater. The system effectively overcomes the limitations of high-temperature and high-pressure digestion methods employed in traditional COD analysis.

View Article and Find Full Text PDF

Ethyl cellulose (EC) is a biocompatible, renewable, and recyclable material with diverse sources, making it an attractive candidate for industrial applications. Electrospinning has gained significant attention for the production of EC fibers. However, conventional electrospinning methods face challenges such as bead formation, low yield, and the absence of porous internal structures, limiting both the functional performance and scalability.

View Article and Find Full Text PDF

Background: During the COVID-19 pandemic, managing respiratory failure in critically ill patients has presented significant challenges. A high-flow nasal cannula (HFNC) has been established as an effective respiratory support modality, offering heated, humidified oxygen at high flow rates. However, concerns persist regarding the potential for aerosol dispersion and the risk of viral transmission, particularly in COVID-19.

View Article and Find Full Text PDF

Introduction: Before performing cell therapy clinical trials, it is important to understand how cells are influenced by different growth conditions and to find optimal xeno-free medium formulations. In this study we have investigated the properties of adipose tissue-derived stem cells (ASCs) cultured under xeno-free conditions.

Methods: Human lipoaspirate samples were digested to yield the stromal vascular fraction cells which were then seeded in i) Minimum Essential Medium-α (MEM-α) supplemented with 10 % (v/v) fetal bovine serum (FBS), ii) MEM-α supplemented with 2 % (v/v) human platelet lysate (PLT) or iii) PRIME-XV MSC expansion XSFM xeno-free, serum free medium (XV).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!