Objective: The objective of this prospective study is to evaluate the consistency of renal stone volume estimation using dual-energy CT across scanner model and reconstruction algorithm configurations.

Subjects And Methods: Patients underwent scanning with routine kidney stone composition protocols on both second- and third-generation dual-source CT scanners. Images were reconstructed using filtered back projection and iterative reconstruction (IR). In addition, a modified IR kernel on the third-generation CT scanner was evaluated. Individual kidney stone volumes were determined and compared.

Results: No significant difference was noted in measured volumes between filtered back-projection data, IR data from the second-generation scanner, and the modified IR kernel data (p > 0.05). The third-generation commercially available IR kernel yielded lower volumes than did the other configurations (p < 0.0001).

Conclusion: With the use of a modified kernel for the third-generation scanner, patients being monitored for changes in kidney stone volume can undergo scanning performed with second- or third-generation dual-energy CT scanners, and the images obtained can be reconstructed with either filtered back projection or IR without the introduction of bias into kidney stone volume measurements.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5481469PMC
http://dx.doi.org/10.2214/AJR.16.16940DOI Listing

Publication Analysis

Top Keywords

stone volume
16
kidney stone
16
modified kernel
12
consistency renal
8
renal stone
8
volume measurements
8
scanner model
8
model reconstruction
8
reconstruction algorithm
8
second- third-generation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!