Background: Due to its great sensitivity, the nucleic acid amplification test (NAAT) is widely used for detection of respiratory viruses (RV). However, few reports have described a direct comparison between multiplex RT-PCR assays for RV. The objective of this study was to perform a direct comparison of three multiplex RT-PCR assays for the detection of respiratory viruses.
Methods: A total of 201 respiratory samples (161 nasopharyngeal swab samples and 40 sputum samples) were tested with three commercial RV assays: Seegene Anyplex II RV16 (AP), LG AdvanSure RV (AD), and Biosewoom Real-Q RV (RQ). The additional tests for the discrepant results were conducted by repeat RV assay or monoplex PCR coupled direct sequencing. Data analysis using percent agreement, kappa, and prevalence-adjusted and bias-adjusted kappa (PABAK) values was performed for comparisons among the three RV assays.
Results: Of the 201 samples, AP, AD, and RQ detected 105 (52.2%), 99 (49.3%), and 95 (47.3%) positive cases respectively. The overall agreement, kappa, and PABAK values for the three assays ranged between 97%-98%, 0.76-0.86, and 0.93-0.96 respectively. The performance of the three assays was very similar, with 94%-100% agreement for all comparisons, each virus types. The additional testing of samples showed discrepant results demonstrating that AD assay had the highest rate of concordance with original results.
Conclusions: We suggest that all multiplex assay would be suitable for the detection of for respiratory viruses in clinical setting.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5836940 | PMC |
http://dx.doi.org/10.1002/jcla.22230 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!