Background: The point-of-care test Roche CARDIAC POC Troponin T (PoC TnT) is an improved assay which has been developed for the Roche cobas h 232 system.

Methods: We performed a multicentre evaluation (four sites) to assess the analytical performance of the PoC TnT assay and to compare it with the central laboratory Elecsys® troponin T high sensitive (lab cTnT-hs) assay.

Results: The relative mean differences found in method comparisons of PoC TnT vs. lab cTnT-hs ranged from -4.1% to +6.8%. Additionally, there was good concordance between PoC TnT and lab cTnT-hs for the number of samples with troponin T values below the measuring range of 40 ng/L. Lot-to-lot differences of PoC TnT ranged from -8.6% to +4.6%. Within-series coefficients of variation (CV) resulting from 81 ten-fold measurements with patient samples were 9.3%, 11.8%, and 12.9% in the low (40 to < 200 ng/L), medium (200 to < 600 ng/L), and high (600 to 2000 ng/L) measuring range, respectively. Using the system quality control, the mean CV for between-day imprecision was 11.3%. No interference was observed by triglycerides (up to 11.4 mmol/L), bilirubin (up to 376 µmol/L), hemoglobin (up to 0.12 mmol/L), biotin (up to 30 µg/L), rheumatoid factor (up to 200 IU/mL), or with 52 standard or cardiovascular drugs at therapeutic concentrations. There was no influence on the results by varying hematocrit values in a range from 25% to 53%. However, interferences with human anti-mouse antibodies were found. No significant influence on the results was found with PoC TnT by using sample volumes between 135 to 165 µL. High troponin T concentrations up to 500 µg/L did not lead to false low results, indicating no high-concentration hook effect. No cross-reactivity was found between the PoC TnT assay and human skeletal troponin T up to 1000 µg/L (< 0.05%). Diagnostic sensitivity and specificity data of a subpopulation (23 patients) of this study are in agreement with results of another large pre-hospital study.

Conclusions: The PoC TnT assay showed good analytical performance with excellent concordance with the calibration and reference laboratory method. It should therefore be suitable for its intended use in point-of-care settings.

Download full-text PDF

Source
http://dx.doi.org/10.7754/Clin.Lab.2016.160814DOI Listing

Publication Analysis

Top Keywords

poc tnt
32
tnt assay
12
lab ctnt-hs
12
poc
9
tnt
8
analytical performance
8
tnt lab
8
measuring range
8
troponin
6
analytical clinical
4

Similar Publications

Objectives: The objective of this study is to evaluate the analytical and diagnostic performance of a high-sensitivity point-of-care (POC) cardiac troponin I assay, the Quidel TriageTrue™ (QuidelOrtho Inc, San Diego, USA), compared to central laboratory testing (CLT) in accelerated diagnostic protocols (ADP) in real time in a clinical environment.

Methods: In a nested sub-study of a pragmatic randomised control trial, consecutive patients with suspected acute coronary syndrome (ACS) and chest pain <12 h duration were randomised to the ESC 0/1 and 0/3-h ADP. Subjects underwent sampling for Quidel TriageTrue POC hs-TnI whole blood and plasma, CLT hs-TnT Roche Elecsys and a validated, NICE approved CLT High sensitivity cardiac troponin I (hs-TnI) (Siemens Attellica) at each time point.

View Article and Find Full Text PDF

Background: Postoperative complication (POC) adversely impacts long-term survival in patients with gastric cancer, perhaps due in part to lower rates for receipt of multimodality therapy (MMT). We sought to determine the impact of POC on MMT completion rates and overall survival (OS) in patients with locally advanced gastric cancer.

Study Design: We analyzed 206 patients with locally advanced gastric cancer undergoing curative-intent resection from 2001 to 2015.

View Article and Find Full Text PDF

Early diagnosis of myocardial infarction (MI) with cardiac troponin (cTn) assays at the point-of-care (POC) is suggested to shorten turn-around-time in the emergency department (ED). The present study aimed at comparing the diagnostic performance of two POC cTn assays with that of a central laboratory high-sensitivity (hs) method, under routine ED conditions. In 2,163 non-selected ED patients suspected for MI, the diagnostic performance of the POC troponin I (TnI), troponin T (TnT), and hs-TnT assay for the prediction of MI was evaluated based on receiver operating characteristic (ROC) analyses and compared with the performance based on the manufacturers' cut-offs.

View Article and Find Full Text PDF

Background: The point-of-care test Roche CARDIAC POC Troponin T (PoC TnT) is an improved assay which has been developed for the Roche cobas h 232 system.

Methods: We performed a multicentre evaluation (four sites) to assess the analytical performance of the PoC TnT assay and to compare it with the central laboratory Elecsys® troponin T high sensitive (lab cTnT-hs) assay.

Results: The relative mean differences found in method comparisons of PoC TnT vs.

View Article and Find Full Text PDF

Background: A multitude of troponin assays for the point-of-care (POC) have been developed showing a lack of analytical sensitivity and precision. We present a new platform solution for the high-sensitivity detection of cardiac troponin T (cTnT) in a 30 μL whole blood sample with a turnaround time of 11 min.

Methods: The immunoassay was completely run in a ready-to-use plastic disposable, a centrifugal microfluidic disc with fully integrated reagents.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!