Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Unlabelled: We investigated the sensitivity of distal bone density, structure, and strength measurements by high-resolution peripheral quantitative computed tomography (HR-pQCT) to variability in limb length. Our results demonstrate that HR-pQCT should be performed at a standard %-of-total-limb-length to avoid substantial measurement bias in population study comparisons and the evaluation of individual skeletal status in a clinical context.
Introduction: High-resolution peripheral quantitative computed tomography (HR-pQCT) measures of bone do not account for anatomic variability in bone length: a 1-cm volume is acquired at a fixed offset from an anatomic landmark. Our goal was to evaluate HR-pQCT measurement variability introduced by imaging fixed vs. proportional volumes and to propose a standard protocol for relative anatomic positioning.
Methods: Double-length (2-cm) scans were acquired in 30 adults. We compared measurements from 1-cm sub-volumes located at the default fixed offset, and the average %-of-length offset. The average position corresponded to 4.0% ± 1.1 mm for radius, and 7.2% ± 2.2 mm for tibia. We calculated the RMS difference in bone parameters and T-scores to determine the measurement variability related to differences in limb length. We used anthropometric ratios to estimate the mean limb length for published HR-pQCT reference data, and then calculated mean %-of-length offsets.
Results: Variability between fixed vs. relative scan positions was highest in the radius, and for cortical bone in general (RMS difference Ct.Th = 19.5%), while individuals had T-score differentials as high as +3.0 SD (radius Ct.BMD). We estimated that average scan position for published HR-pQCT reference data corresponded to 4.0% at the radius, and 7.3% at tibia.
Conclusion: Variability in limb length introduces significant bias to HR-pQCT measures, confounding cross-sectional analyses and limiting the clinical application for individual assessment of skeletal status. We propose to standardize scan positioning using 4.0 and 7.3% of total bone length for the distal radius and tibia, respectively.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5526099 | PMC |
http://dx.doi.org/10.1007/s00198-017-4010-7 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!