Myopic Laser Corneal Refractive Surgery Reduces Interdevice Agreement in the Measurement of Anterior Corneal Curvature.

Eye Contact Lens

Department of Ophthalmology (H.J., P.Z.), Xinhua Hospital Affiliated to Medical College of Shanghai Jiaotong University, Shanghai, China; Department of Physics (Z.O.), Shanghai University, Shanghai, China; and Department of Ophthalmology (H.G.), Aier Eye Hospital, Zhengzhou, China.

Published: September 2018

Objectives: To investigate interdevice differences and agreement in the measurement of anterior corneal curvature obtained by different technologies after laser corneal refractive surgery.

Methods: The prospective study comprised 109 eyes of 109 consecutive patients who had undergone laser-assisted in situ keratomileusis (LASIK). Preoperative and postoperative corneal parameters were measured by Scheimpflug imaging (Pentacam), Placido-slit-scanning (Orbscan) and auto-keratometry (IOLMaster). Preoperative and postoperative anterior corneal curvatures (K readings) were compared between devices. Interdevice agreement was evaluated by Bland-Altman analysis.

Results: Preoperatively, the difference of K reading for Pentacam-IOLMaster (0.04±0.20 D) was not statistically significant (P=0.059). The differences between Pentacam-Orbscan and Orbscan-IOLMaster were 0.20±0.34 D (P<0.001) and -0.17±0.29 D (P<0.001), respectively. After surgery, no difference was found for Pentacam-Orbscan (-0.05±0.38, P=0.136). The differences between Pentacam-IOLMaster and Orbscan-IOLMaster were 0.13±0.29 D (P<0.001) and 0.19±0.34 D (P<0.001). Preoperative interdevice agreement (95% limit of agreement [LOA]) between Pentacam and Orbscan, Pentacam and IOLMaster, and Orbscan and IOLMaster were 1.31 D, 0.79 D and 1.14 D, respectively. The 95% LOAs decreased to 1.47 D, 1.14 D, and 1.34 D after refractive surgery.

Conclusion: Corneal refractive surgery changed the preoperative and postoperative interdevice differences in corneal curvature measurements and reduced interdevice agreement, indicating that the devices are not interchangeable.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ICL.0000000000000364DOI Listing

Publication Analysis

Top Keywords

anterior corneal
12
laser corneal
8
corneal refractive
8
interdevice agreement
8
agreement measurement
8
measurement anterior
8
corneal curvature
8
preoperative postoperative
8
corneal
6
myopic laser
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!