  • Cervical total disc replacement aims to maintain motion and prevent degeneration but may risk altered facet joint loading and arthritis over time.
  • The study used fresh-frozen human cadaver spines to measure changes in facet joint forces and pressures during different surgical interventions, including disc replacement and fusion.
  • Findings revealed that while disc replacement preserved facet contact force under most conditions, it resulted in increased force under extension, which might increase the risk of joint disease at the treated level.

Article Abstract

Background: Cervical total disc replacement was developed to preserve motion and reduce adjacent-level degeneration relative to fusion, yet concerns remain that total disc replacement will lead to altered facet joint loading and long-term facet joint arthrosis. This study is intended to evaluate changes in facet contact force, pressure and surface area at the treated and superior adjacent levels before and after discectomy, disc replacement, and fusion.

Methods: Ten fresh-frozen human cadaveric cervical spines were potted from C2 to C7 with pressure sensors placed into the facet joints of C3-C4 and C4-C5 via slits in the facet capsules. Moments were applied to the specimens to produce axial rotation, lateral bending and extension. Facet contact force and pressure were measured at both levels for intact, discectomy at C4-C5, disc replacement with ProDisc-C (Synthes Spine, West Chester, Pennsylvania, USA) at C4-C5, and anterior discectomy and fusion with Cervical Spine Locking Plate (Synthes Spine, West Chester, Pennsylvania, USA) at C4-C5. Facet contact area was calculated from the force and pressure measurements. An analysis of variance was used to determine significant differences with P-values <0.05 indicating significance.

Findings: Facet contact force was elevated at the treated level under extension following both discectomy and disc replacement, while facet contact pressure and area were relatively unchanged. Facet contact force and area were decreased at the treated level following fusion for all three loading conditions.

Interpretation: Total disc replacement preserved facet contact force for all scenarios except extension at the treated level, highlighting the importance of the anterior disco-ligamentous complex. This could promote treated-level facet joint disease.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiomech.2017.03.007DOI Listing

Publication Analysis

Top Keywords

disc replacement
20
facet contact
12
force pressure
12
total disc
8
facet joint
8
contact force
8
synthes spine
8
spine west
8
west chester
8
chester pennsylvania
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!