  • The study investigates the eligibility of the subcutaneous implantable cardioverter-defibrillator (S-ICD) in patients with inherited arrhythmia syndromes, focusing particularly on those with Brugada syndrome (BrS).
  • Researchers screened 100 patients with various arrhythmia syndromes, finding that 56% of BrS patients had spontaneous type 1 ECG, while others required a drug (ajmaline) to reveal the condition.
  • Overall, 13% of patients faced screening failures, with BrS patients experiencing a higher failure rate and fewer suitable sensing vectors compared to patients with other channelopathies.

Article Abstract

Aims: Subcutaneous implantable cardioverter-defibrillator (S-ICD) can avoid important complications associated with transvenous leads in patients with inherited primary arrhythmia syndromes, who do not need pacing therapy. Few data are available on the percentage of patients with inherited arrhythmia syndromes eligible for S-ICD implantation. Aim of this study was to analyse the eligibility for S-ICD in a series of patients with Brugada syndrome (BrS), and to compare it with patients with other channelopathies.

Methods And Results: Patients presenting with BrS, long-QT syndrome (LQTS), early repolarization syndrome (ERS), and idiopathic ventricular fibrillation (IVF) were considered eligible for this study. ECG screening was performed by analysis of QRS complex and T wave morphology recorded in standing and supine position. Eligibility was defined when ≥1 sense vector was acceptable in both supine and standing position. A total of 100 patients (72 males; mean age: 46 ± 17 years) underwent S-ICD sensing screening. Sixty-one patients presented with BrS, 21 with LQTS, 14 with IVF, and 4 with ERS. Thirty-four patients with BrS (56%) presented with spontaneous type 1 ECG. In the other 27 patients (44%), type 1 ECG was unmasked by ajmaline. Overall, rate of screening failure was 13%. Patients with BrS had a higher rate of inappropriate morphology analysis as compared with other channelopathies (18% vs. 5%, P = 0.07) and had a lower number of suitable sensing vectors (49.6% vs. 84.7% vs. P < 0.001). Ajmaline challenge unmasked sensing failure in 14.8% of drug-induced BrS patients previously considered eligible. In all patients, the reason for sensing inappropriateness was due to the presence of high T wave voltages.

Conclusion: S-ICD screening failure occurs in up to 13% of patients with inherited primary arrhythmia syndromes. Patients with BrS present a higher rate of screening failure as compared with other cardiac channelopathies.

Download full-text PDF

Source
http://dx.doi.org/10.1093/europace/eux009DOI Listing

Publication Analysis

Top Keywords

arrhythmia syndromes
12
patients
11
subcutaneous implantable
8
implantable cardioverter-defibrillator
8
sensing screening
8
screening failure
8
patients brugada
8
brugada syndrome
8
inherited primary
8
primary arrhythmia
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!