Purpose: Onyx has been successfully applied in the treatment of various neurovascular lesions. However, some experience is required to get accustomed to its unpredictable fluoroscopic visibility during injection. This study aimed to evaluate the characteristics of radiopacity change in a simulated embolization procedure.
Materials And Methods: Using a bench-top Onyx injection experiment simulating a typical brain arteriovenous malformation embolization, nine cycles of casting modes (continuous injection) and plugging modes (injection with intermittent pauses) were performed. Radiodensity of Onyx droplets collected from the microcatheter tip and the distal head portion of the microcatheter were measured as time lapsed. Distribution of droplet radiodensity (radiodensity) and distribution of radiographic grade (grade) were analyzed and compared by repeated measurements.
Results: Within-group analysis revealed no significant radiodensity change with time (P>0.05). The radiodensity was significantly higher in the casting mode than in the plugging mode (P<0.01). The lateral radiograph of the microcatheter showed higher radiopacity (P<0.01) and better evenness (P<0.01) in the casting mode than in the plugging mode. A significant difference in microcatheter attenuation (both radiographic grade mean and SD; P<0.01) was noted between the two modes. Radiodensity had a significant influence on the radiopacity and radiopacity evenness of the microcatheter.
Conclusion: The radiopacity of the Onyx can vary significantly over time because of early precipitation of tantalum powder. Radiopacity decreased significantly during plugging modes, characterized by pauses between injections.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5355458 | PMC |
http://dx.doi.org/10.5469/neuroint.2017.12.1.3 | DOI Listing |
Sci Rep
December 2024
Postgraduate Program in Health Sciences, Health Sciences Center, Federal University of Rio Grande do Norte, Natal, RN, Brazil.
Body composition abnormalities are prognostic markers in several types of cancer, including colorectal cancer (CRC). Using our data distribution on body composition assessments and classifications could improve clinical evaluations and support population-specific opportune interventions. This study aimed to evaluate the distribution of body composition from computed tomography and assess the associations with overall survival among patients with CRC.
View Article and Find Full Text PDFJ Cachexia Sarcopenia Muscle
February 2025
Clinical Surgery, University of Edinburgh, Royal Infirmary of Edinburgh, Edinburgh, Scotland, UK.
J Cachexia Sarcopenia Muscle
February 2025
Department of Internal Medicine, Division of Medical Oncology, Yonsei Cancer Center, Yonsei University College of Medicine, Seoul, South Korea.
Background: Recent evidence indicates that a dysregulated host metabolism influences treatment outcomes in patients with breast cancer. We investigated the association of computed tomography (CT)-derived body composition indices with therapeutic responses in patients with hormone receptor-positive, HER2-negative advanced breast cancer (ABC) on endocrine plus CDK4/6 inhibitor (CDK4/6i) treatment.
Methods: The study involved a retrospective cohort of patients with ABC at the Yonsei Cancer Center who received CDK4/6i and aromatase inhibitors as first-line therapy between January 2017 and October 2020.
J Biomech
December 2024
Clemson University, Department of Bioengineering, 301 Rhodes Building, Clemson, SC 29634, United States; Aravis BioTech LLC, 17 Claret Dr., Greenville SC 29609, United States. Electronic address:
Bone healing after sliding hip screw internal fixation of intertrochanteric hip fractures is difficult to monitor with radiography. In this study, we describe and evaluate a device to non-invasively determine the loading on the screw implant as a possible qualitative indicator of bone healing. A novel load-sensing sliding hip screw (LS-SHS) was fabricated containing a radio-dense tungsten indicator rod that moves and can be measured within the screw cannulation when the screw bends under load via plain radiography.
View Article and Find Full Text PDFJ Orthop Case Rep
December 2024
Department of Orthopaedics, AIIMS Raebareli, Raebareli, Uttar Pradesh, India.
Introduction: Metallosis, characterized by the collection of metallic debris in periarticular tissues, is primarily associated with metal-on-metal bearings in hip arthroplasty. This report presents a rare case of metallosis after metal-on-polyethylene (MoP) total hip arthroplasty (THA), highlighting diagnostic challenges and management strategies. Metallosis following MoP arthroplasty is extremely rare, with only a limited number of documented cases in the literature.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!