A stable and convenient protein electrophoresis titration device with bubble removing system.

Electrophoresis

Laboratory of Bioseparation and Analytical Biochemistry, State Key Laboratory of Microbial Metabolism, School of Life Science and Biotechnology, Shanghai Jiao Tong University, Shanghai, P.R. China.

Published: July 2017

Moving reaction boundary titration (MRBT) has a potential application to immunoassay and protein content analysis with high selectivity. However, air bubbles often impair the accuracy of MRBT, and the leakage of electrolyte greatly decreases the safety and convenience of electrophoretic titration. Addressing these two issues a reliable MRBT device with modified electrolyte chamber of protein titration was designed. Multiphysics computer simulation was conducted for optimization according to two-phase flow. The single chamber was made of two perpendicular cylinders with different diameters. After placing electrophoretic tube, the resident air in the junction next to the gel could be eliminated by a simple fast electrolyte flow. Removing the electrophoretic tube automatically prevented electrolyte leakage at the junction due to the gravity-induced negative pressure within the chamber. Moreover, the numerical simulation and experiments showed that the improved MRBT device has following advantages: (i) easy and rapid setup of electrophoretic tube within 20 s; (ii) simple and quick bubble dissipates from the chamber of titration within 2 s; (iii) no electrolyte leakage from the two chambers: and (iv) accurate protein titration and safe instrumental operation. The developed technique and apparatus greatly improves the performance of the previous MRBT device, and providing a new route toward practical application.

Download full-text PDF

Source
http://dx.doi.org/10.1002/elps.201600472DOI Listing

Publication Analysis

Top Keywords

mrbt device
12
electrophoretic tube
12
protein titration
8
electrolyte leakage
8
titration
6
mrbt
5
electrolyte
5
stable convenient
4
protein
4
convenient protein
4

Similar Publications

Objectives: Patent hemostasis (PH) is essential for preventing radial artery occlusion (RAO) after trans-radial procedures; however, it remains unclear how it should be obtained. The aim of this multicenter randomized study was to evaluate whether the use of an adjustable device (AD), inflated with a pre-determined amount of air (AoA), was more effective than a non-adjustable device (non-AD) for achieving PH, thereby reducing the incidence of RAO.

Methods: We enrolled a total of 480 patients undergoing transradial procedure at 3 Italian institutions.

View Article and Find Full Text PDF

A stable and convenient protein electrophoresis titration device with bubble removing system.

Electrophoresis

July 2017

Laboratory of Bioseparation and Analytical Biochemistry, State Key Laboratory of Microbial Metabolism, School of Life Science and Biotechnology, Shanghai Jiao Tong University, Shanghai, P.R. China.

Moving reaction boundary titration (MRBT) has a potential application to immunoassay and protein content analysis with high selectivity. However, air bubbles often impair the accuracy of MRBT, and the leakage of electrolyte greatly decreases the safety and convenience of electrophoretic titration. Addressing these two issues a reliable MRBT device with modified electrolyte chamber of protein titration was designed.

View Article and Find Full Text PDF

Fast and selective determination of total protein in milk powder via titration of moving reaction boundary electrophoresis.

Electrophoresis

May 2013

Laboratory of Bioseparation and Analytical Biochemistry, State Key Laboratory of Microbial Metabolism, School of Life Science and Biotechnology, Shanghai Jiao Tong University, Shanghai, China.

In this paper, moving reaction boundary titration (MRBT) was developed for rapid and accurate quantification of total protein in infant milk powder, from the concept of moving reaction boundary (MRB) electrophoresis. In the method, the MRB was formed by the hydroxide ions and the acidic residues of milk proteins immobilized via cross-linked polyacrylamide gel (PAG), an acid-base indicator was used to denote the boundary motion. As a proof of concept, we chose five brands of infant milk powders to study the feasibility of MRBT method.

View Article and Find Full Text PDF