Objectives: To evaluate the accuracy of reduced-dose CT scans reconstructed using a new generation of model-based iterative reconstruction (MBIR) in the imaging of urinary tract stone disease, compared with a standard-dose CT using 30% adaptive statistical iterative reconstruction.
Methods: This single-institution prospective study recruited 125 patients presenting either with acute renal colic or for follow-up of known urinary tract stones. They underwent two immediately consecutive scans, one at standard dose settings and one at the lowest dose (highest noise index) the scanner would allow. The reduced-dose scans were reconstructed using both ASIR 30% and MBIR algorithms and reviewed independently by two radiologists. Objective and subjective image quality measures as well as diagnostic data were obtained.
Results: The reduced-dose MBIR scan was 100% concordant with the reference standard for the assessment of ureteric stones. It was extremely accurate at identifying calculi of 3 mm and above. The algorithm allowed a dose reduction of 58% without any loss of scan quality.
Conclusions: A reduced-dose CT scan using MBIR is accurate in acute imaging for renal colic symptoms and for urolithiasis follow-up and allows a significant reduction in dose.
Key Points: • MBIR allows reduced CT dose with similar diagnostic accuracy • MBIR outperforms ASIR when used for the reconstruction of reduced-dose scans • MBIR can be used to accurately assess stones 3 mm and above.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00330-017-4783-2 | DOI Listing |
Acta Orthop
January 2025
Emeritus Consultant Orthopaedic Surgeon, Wrightington Hospital; Bristol University, UK.
Background And Purpose: The amount of information publicly available from arthroplasty registries is large but could be used more effectively. This project aims to improve the knowledge concerning existing registries to facilitate access, transparency, harmonization, and reporting.
Methods: Within the International Society of Arthroplasty Registries (ISAR) we aimed at developing, testing, adopting, and making publicly available a short, standardized registry description with items considered relevant for stakeholders using a cross-sectional study survey.
Neuroradiol J
January 2025
Department of Radiology and Radiological Science, Medical University of South Carolina, USA.
We describe a novel application of photon-counting detector CT (PCD-CT) in neurovascular imaging by harnessing the improved spatial resolution, attenuation of electronic noise, and reduction of metal artifacts. The presented case offers the unique challenge of high-quality imaging for the assessment of treated and untreated intracranial saccular aneurysms, in the setting of metal artifacts from embolization coils. Our goal was to explore optimized reconstruction parameters for ultra-high-resolution imaging (UHR) using a dedicated, sharp neurovascular kernel (Hv72) and the highest strength of quantum iterative reconstruction (QIR-4) for detailed characterization of the vasculature.
View Article and Find Full Text PDFJ Comput Assist Tomogr
January 2025
Department of Radiology, the Affiliated Traditional Chinese Medicine Hospital, Southwest Medical University.
Background: With the widespread use of lumbar pedicle screws for internal fixation, the morphology of the screws and the surrounding tissues should be evaluated. The metal artifact reduction (MAR) technique can reduce the artifacts caused by pedicle screws, improve the quality of computed tomography (CT) images after pedicle fixation, and provide more imaging information to the clinic.
Purpose: To explore whether the MAR+ method, a projection-based algorithm for correcting metal artifacts through multiple iterative operations, can reduce metal artifacts and have an impact on the structure of the surrounding metal.
Synchrotron X-ray microtomography (S-µCT) is a highly valuable technique for investigating organ function and pathologies. However, its application is often limited by high radiation doses and the occurrence of ring artifacts. While S-µCT utilizing sparse-view projections can effectively decrease radiation doses, the reconstructed images frequently exhibit severe streaking artifacts, which are exacerbated by ring artifacts, ultimately compromising reconstruction accuracy, image quality, and resolution.
View Article and Find Full Text PDFDue to their advantages of compact geometries and lightweight, diffractive optical elements (DOEs) are attractive in various applications such as sensing, imaging and holographic display. When designing DOEs based on algorithms, a diffraction model is required to trace the diffracted light propagation and to predict the performance. To have more precise diffraction field tracing and optical performance simulation, different diffraction models have been proposed and developed.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!