Introduction: The range shifter (RS) is used to treat shallow tumors for a proton pencil beam scanning system (PBS). Adding RS certainly complicates the commissioning of the treatment planning system (TPS) because the spot sizes are significantly enlarged with RS. In this work, we present an efficient method to configure a commercial TPS for a PBS system with a fixed RS.

Methods: By combining a spiral delivery with customized control points, we were able to significantly improve measurement efficiency and obtain 250 field size factors (FSF) within three hours. The measured FSFs were used to characterize the proton fluence and fit the parameters for the double-Gaussian fluence model used in the TPS. Extensive validation was performed using FSFs measured in air and in water, absolute doses of spread-out Bragg peak (SOBP) fields, and the dose measurements carried out for patient-specific quality assurance (QA).

Results: The measured in-air FSFs agreed with the model's prediction within 3% for all 250 FSFs, and within 2 for 94% of the FSFs. The agreement between model's prediction and measurement was within 2% for the in-air and in-water FSFs and the absolute doses for SOBP beams. The patient-specific QA of 113 fields showed an excellent gamma passing rates (96.95 ± 2.51%) for the absolute dose comparisons with gamma criteria of 2 mm and 2%.

Conclusion: The excellent agreement between the model's prediction and measurements proved the efficiency and accuracy of the proposed method of using FSFs to characterize the proton fluence and configure the TPS for a PBS system with fixed RS.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5348744PMC
http://dx.doi.org/10.1186/s13014-017-0783-2DOI Listing

Publication Analysis

Top Keywords

model's prediction
12
field size
8
size factors
8
range shifter
8
tps pbs
8
pbs system
8
system fixed
8
fsfs characterize
8
characterize proton
8
proton fluence
8

Similar Publications

Comprehensive Analysis Reveals the Potential Diagnostic Value of Biomarkers Associated With Aging and Circadian Rhythm in Knee Osteoarthritis.

Orthop Surg

January 2025

Department of Orthopedics, Tianjin Medical University General Hospital, International Science and Technology Cooperation Base of Spinal Cord Injury, Tianjin Key Laboratory of Spine and Spinal Cord, Tianjin, China.

Objective: Knee osteoarthritis (KOA) is characterized by structural changes. Aging is a major risk factor for KOA. Therefore, the objective of this study was to examine the role of genes related to aging and circadian rhythms in KOA.

View Article and Find Full Text PDF

Impeding linear calibration models from accurately predicting target sample analyte amounts are the target sample-wise deviations in measurement profiles (e.g., spectra) relative to calibration samples.

View Article and Find Full Text PDF

Zero echo time (zero-TE) pulse sequences provide a quiet and artifact-free alternative to conventional functional magnetic resonance imaging (fMRI) pulse sequences. The fast readouts (<1 ms) utilized in zero-TE fMRI produce an image contrast with negligible contributions from blood oxygenation level-dependent (BOLD) mechanisms, yet the zero-TE contrast is highly sensitive to brain function. However, the precise relationship between the zero-TE contrast and neuronal activity has not been determined.

View Article and Find Full Text PDF

Background: Established risk models may not be applicable to patients at higher cardiovascular risk with a measured Lp(a) (lipoprotein[a]) level, a causal risk factor for atherosclerotic cardiovascular disease.

Methods: This was a model development study. The data source was the Nashville Biosciences Lp(a) data set, which includes clinical data from the Vanderbilt University Health System.

View Article and Find Full Text PDF

Semiparametric estimator for the covariate-specific receiver operating characteristic curve.

Stat Methods Med Res

January 2025

CITMAga and Department of Statistics and Operations Research, Universidade de Vigo, Vigo, Galicia, Spain.

The study of the predictive ability of a marker is mainly based on the accuracy measures provided by the so-called confusion matrix. Besides, the area under the receiver operating characteristic curve has become a popular index for summarizing the overall accuracy of a marker. However, the nature of the relationship between the marker and the outcome, and the role that potential confounders play in this relationship could be fundamental in order to extrapolate the observed results.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!