Validation of Noninvasive Measurement of Cardiac Output Using Inert Gas Rebreathing in a Cohort of Patients With Heart Failure and Reduced Ejection Fraction.

Circ Heart Fail

From the Cardiology Department (M.H., A.E., M.F.E.) and Radiology Department (A.K.), Cairo University, Egypt; Cardiology Department (M.H., K.W., P.P.S., A.E., H.M.), Radiology Department (A.K.), Biomedical Engineering Department (A.N.), and Cardiothoracic Department (M.H.Y.), Aswan Heart Centre, Egypt; and Imperial College, London, United Kingdom (M.H.Y.).

Published: March 2017

Background: Cardiac output (CO) is a key indicator of cardiac function in patients with heart failure. No completely accurate method is available for measuring CO in all patients. The objective of this study was to validate CO measurement using the inert gas rebreathing (IGR) method against other noninvasive and invasive methods of CO quantification in a cohort of patients with heart failure and reduced ejection fraction.

Methods And Results: The study included 97 patients with heart failure and reduced ejection fraction (age 42±15.5 years; 64 patients (65.9%) had idiopathic dilated cardiomyopathy and 21 patients (21.6%) had ischemic heart disease). Median left ventricle ejection fraction was 24% (10%-36%). Patients with atrial fibrillation were excluded. CO was measured using 4 methods (IGR, cardiac magnetic resonance imaging, cardiac catheterization, and echocardiography) and indexed to body surface area (cardiac index [CI]). All studies were performed within 48 hours. Median CI measured by IGR was 1.75, by cardiac magnetic resonance imaging was 1.82, by cardiac catheterization was 1.65, and by echo was 1.7 L·min·m. There were significant modest linear correlations between IGR-derived CI and cardiac magnetic resonance imaging-derived CI (=0.7; <0.001), as well as cardiac catheterization-derived CI (=0.6; <0.001). Using Bland-Altman analysis, the agreement between the IGR method and the other methods was as good as the agreement between any 2 other methods with each other.

Conclusions: The IGR method is a simple, accurate, and reproducible noninvasive method for quantification of CO in patients with advanced heart failure. The prognostic value of this simple measurement needs to be studied prospectively.

Download full-text PDF

Source
http://dx.doi.org/10.1161/CIRCHEARTFAILURE.116.003592DOI Listing

Publication Analysis

Top Keywords

patients heart
16
heart failure
16
failure reduced
12
reduced ejection
12
ejection fraction
12
cardiac magnetic
12
magnetic resonance
12
cardiac
9
cardiac output
8
inert gas
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!