Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: We evaluated the effect of changes in FiO on the bias and accuracy of the determination of oxygen consumption (V˙O) and carbon dioxide production (V˙CO) using the E-COVX monitor in patients with mechanical ventilation.
Design: Descriptive of concordance.
Setting: Intensive Care Unit.
Patients Or Participants: Patients with mechanical ventilation.
Interventions: We measured V˙O and V˙CO using the E-COVX monitor. Values recorded were the average in 5min. Two groups of 30 patients. We analyzed: 1) the reproducibility in the measurement of V˙O and V˙CO at FiO 0.4, and 2) the effect of the changes in FiO on the measurement of V˙O and V˙CO. Statistical analysis was performed using Bland and Altman test.
Variables Of Main Interest: Bias and accuracy.
Results: 1) FiO 0.4 reproducibility: The bias in the measurement of V˙O and V˙CO was 1.6 and 2.1mL/min, respectively, and accuracy was 9.7 to -8.3% and 7.2 to -5.2%, respectively, and 2) effect of FiO on V˙O: The bias of V˙O measured at FiO 0.4 and 0.6 was -4.0mL/min and FiO 0.4 and 0.8 was 5.2mL/min. Accuracy between FiO 0.4 and 0.6 was 11.9 to -14.1%, and between FiO 0.4 and 0.8 was 43.9 to -39.7%.
Conclusions: The E-COVX monitor evaluates V˙O and V˙CO in critical patients with mechanical ventilation with a clinically acceptable accuracy until FiO 0.6.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.medin.2016.12.002 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!