Objectives: The aim of this study is to evaluate the marginal and internal fit of zirconia core crowns manufactured following different digital and traditional workflows.

Methods: A 6° taper shoulder prepared abutment tooth was used to produce 20 zirconia core crowns using four different scanning techniques: scanned directly with the extraoral lab scanner, scanned with intraoral scanner, dental impressions using individual dental tray and polyether, dental casts from a polyether impressions. Marginal and internal fits were evaluated with digital photography and the silicone replica method.

Results: Medium marginal gaps were 76,00 μm ± 28.9 for extraoral lab scanner, 80.50 μm ± 36,2 for intraoral scanner, 88.10 μm ± 34,8 for dental impression scan and 112,4 μm ± 37,2 for dental cast scan. Medium internal gaps were 23.20 μm ± 10,3 for extraoral lab scanner, 16.20 μm ± 8.3 for intraoral scanner, 27.20 μm ± 16.7 for dental impression scan and 30.20 μm ± 12.7 for dental cast scan.

Conclusion: Internal gap were extensively lower than 70 μm described in literature. Marginal fit was higher than ideal values for all the techniques but within the limit of clinical success. Intraoral scanners obtained the best results for internal gap.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5333748PMC
http://dx.doi.org/10.11138/orl/2016.9.1S.021DOI Listing

Publication Analysis

Top Keywords

extraoral lab
12
lab scanner
12
intraoral scanner
12
μm
9
marginal internal
8
zirconia core
8
core crowns
8
dental impression
8
impression scan
8
dental cast
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!