Background: Several different defibrillators are currently used for cardioversion and defibrillation of cardiac arrhythmias. The efficacy of a novel pulsed biphasic (PB) waveform has not been compared to other biphasic waveforms. Accordingly, this study aims to compare the efficacy and safety of PB shocks with biphasic truncated exponential (BTE) shocks in patients undergoing cardioversion of atrial fibrillation or -flutter.

Methods And Results: This prospective, randomized study included patients admitted for elective direct current cardioversion. Patients were randomized to receive cardioversion using either PB or BTE shocks. We used escalating shocks until sinus rhythm was obtained or to a maximum of 4 shocks. Patients randomized to PB shocks received 90, 120, 150, and 200 J and patients randomized to BTE shocks received 100, 150, 200, and 250 J, as recommended by the manufacturers. In total, 69 patients (51%) received PB shocks and 65 patients (49%) BTE shocks. Successful cardioversion, defined as sinus rhythm 4 hours after cardioversion, was achieved in 43 patients (62%) using PB shocks and in 56 patients (86%) using BTE shocks; ratio 1.4 (95% CI 1.1-1.7) (=0.002). There was no difference in safety (ie, myocardial injury judged by changes in high-sensitive troponin I levels; ratio 1.1) (95% CI 1.0-1.3), =0.15. The study was terminated prematurely because of an adverse event.

Conclusions: Cardioversion using a BTE waveform was more effective when compared with a PB waveform. There was no difference in safety between the 2 waveforms, as judged by changes in troponin I levels.

Clinical Trial Registration: URL: http://www.clinicaltrials.gov. Unique identifier: NCT02317029.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5524016PMC
http://dx.doi.org/10.1161/JAHA.116.004853DOI Listing

Publication Analysis

Top Keywords

bte shocks
20
shocks patients
16
patients randomized
12
shocks
11
patients
9
cardioversion
8
pulsed biphasic
8
biphasic truncated
8
truncated exponential
8
cardioversion bte
8

Similar Publications

Article Synopsis
  • Atrial fibrillation (AF) is a common heart rhythm disorder, and cardioversion is a method used to return the heart to its normal rhythm through medications or electrical shocks.
  • The study aimed to evaluate the effectiveness and safety of both pharmacological and electrical cardioversion treatments for AF and related conditions.
  • Researchers analyzed data from 112 randomized controlled trials involving nearly 16,000 patients, noting that many trials had a high risk of bias and varying demographics among participants.
View Article and Find Full Text PDF
Article Synopsis
  • Bispecific T-cell engagers (BTEs) are new treatments for blood cancers, but past studies didn't effectively explore their cardiovascular side effects.
  • Using data from the FDA's Adverse Event Reporting System, researchers analyzed cardiovascular adverse events (CVAE) related to five FDA-approved BTEs, discovering a notable incidence and risk of these events.
  • The study found that about 20.4% of BTE-related reports involved CVAEs, with significant mortality associated; teclistamab showed the highest risk for severe cardiovascular issues, while blinatumomab was linked to other serious complications.
View Article and Find Full Text PDF

Introduction: The wearable cardioverter defibrillator (WCD) is used to protect patients at risk for sudden cardiac arrest. We examined defibrillation efficacy and safety of a biphasic truncated exponential waveform designed for use in a contemporary WCD in three animal studies and a human study.

Methods: Animal (swine) studies: #1: Efficacy comparison of a 170J BTE waveform (SHOCK A) to a 150J BTE waveform (SHOCK B) that approximates another commercially available waveform.

View Article and Find Full Text PDF

Smaller electrodes allow more options for design of automated external defibrillator (AED) user interfaces. However, previous studies employing monophasic-waveform defibrillators found that smaller electrode sizes have lower defibrillation shock success rates. We hypothesize that, for impedance-compensated, biphasic truncated exponential (BTE) shocks, smaller electrodes increase transthoracic impedance (TTI) but do not adversely affect defibrillation success rates.

View Article and Find Full Text PDF

Aims: The Prehospital Optimal Shock Energy for Defibrillation (POSED) study will assess the feasibility of conducting a cluster randomised controlled study of clinical effectiveness in UK ambulance services to identify the optimal shock energy for defibrillation.

Methods: POSED is a pragmatic, allocation concealed, open label, cluster randomised, controlled feasibility study. Defibrillators within a single UK ambulance service will be randomised in an equal ratio to deliver one of three shock strategies 120-150-200 J, 150-200-200 J, 200-200-200 J.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!