Objective: To investigate the effect of dedicated Computer Aided Detection (CAD) software for automated breast ultrasound (ABUS) on the performance of radiologists screening for breast cancer.

Methods: 90 ABUS views of 90 patients were randomly selected from a multi-institutional archive of cases collected between 2010 and 2013. This dataset included normal cases (n=40) with >1year of follow up, benign (n=30) lesions that were either biopsied or remained stable, and malignant lesions (n=20). Six readers evaluated all cases with and without CAD in two sessions. CAD-software included conventional CAD-marks and an intelligent minimum intensity projection of the breast tissue. Readers reported using a likelihood-of-malignancy scale from 0 to 100. Alternative free-response ROC analysis was used to measure the performance.

Results: Without CAD, the average area-under-the-curve (AUC) of the readers was 0.77 and significantly improved with CAD to 0.84 (p=0.001). Sensitivity of all readers improved (range 5.2-10.6%) by using CAD but specificity decreased in four out of six readers (range 1.4-5.7%). No significant difference was observed in the AUC between experienced radiologists and residents both with and without CAD.

Conclusions: Dedicated CAD-software for ABUS has the potential to improve the cancer detection rates of radiologists screening for breast cancer.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2017.01.021DOI Listing

Publication Analysis

Top Keywords

cancer detection
8
automated breast
8
breast ultrasound
8
computer aided
8
aided detection
8
radiologists screening
8
screening breast
8
breast
5
cad
5
readers
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!