Objective: To evaluate the effectiveness and clinical utility of a metal artifact reduction (MAR) image reconstruction algorithm for the reduction of high-attenuation object (HAO)-related image artifacts.
Methods: Images were quantitatively evaluated for image noise (noiseSD and noiserange) and qualitatively for artifact severity, gray-white-matter delineation, and diagnostic confidence with conventional reconstruction and after applying a MAR algorithm.
Results: Metal artifact reduction reduces noiseSD and noiserange (median [interquartile range]) at the level of HAO in 1-cm distance compared with conventional reconstruction (noiseSD: 60.0 [71.4] vs 12.8 [16.1] and noiserange: 262.0 [236.8] vs 72.0 [28.3]; P < 0.0001). Artifact severity (reader 1 [mean ± SD]: 1.1 ± 0.6 vs 2.4 ± 0.5, reader 2: 0.8 ± 0.6 vs 2.0 ± 0.4) at level of HAO and diagnostic confidence (reader 1: 1.6 ± 0.7 vs 2.6 ± 0.5, reader 2: 1.0 ± 0.6 vs 2.3 ± 0.7) significantly improved with MAR (P < 0.0001). Metal artifact reduction did not affect gray-white-matter delineation.
Conclusions: Metal artifact reduction effectively reduces image artifacts caused by HAO and significantly improves diagnostic confidence without worsening gray-white-matter delineation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/RCT.0000000000000575 | DOI Listing |
J Magn Reson
December 2024
Univ. Lille, CNRS, Centrale Lille, Univ. Artois, UMR 8181, UCCS - Unité de Catalyse et Chimie du Solide, F-59000 Lille, France. Electronic address:
The two-dimensional (2D) refocused INADEQUATE NMR experiment, which correlates double-quantum (DQ) and single-quantum (SQ) coherences, is widely used to probe the chemical connectivities in solids. Nevertheless, the multiplets along the F dimension reduce the resolution and sensitivity of this experiment. The Composite-Refocusing (CR) technique with two excitation pulses has been proposed to suppress these multiplets in 2D INADEQUATE spectra of liquids.
View Article and Find Full Text PDFAdv Radiat Oncol
February 2025
Department of Radiation Oncology, University of Utah, Salt Lake City, Utah.
Purpose: To evaluate the image quality of an ultrafast cone-beam computed tomography (CBCT) system-Varian HyperSight.
Methods And Materials: In this evaluation, 5 studies were performed to assess the image quality of HyperSight CBCT. First, a HyperSight CBCT image quality evaluation was performed and compared with Siemens simulation-CT and Varian TrueBeam CBCT.
LC-ESI-MS/MS is a preferred method for detecting and identifying metabolites, including those that are unpredictable from the genome, especially in basal metazoans like Cnidaria, which diverged earlier than bilaterians and whose metabolism is poorly understood. However, the unexpected appearance of a "ghost peak" for dopamine, which exhibited the same m/z value and MS/MS product ion spectrum during an analysis of Nematostella vectensis, a model cnidarian, complicated its accurate identification. Understanding the mechanism by which "ghost peaks" appear is crucial to accurately identify the monoamine repertoire in early animals so as to avoid misassignments.
View Article and Find Full Text PDFMed Image Anal
December 2024
School of Physics, Beihang University, Beijing, China; Hangzhou International Innovation Institute, Beihang University, Hangzhou, China; Tianmushan Laboratory, Hangzhou, China. Electronic address:
Computed tomography (CT) is continuously becoming a valuable diagnostic technique in clinical practice. However, the radiation dose exposure in the CT scanning process is a public health concern. Within medical diagnoses, mitigating the radiation risk to patients can be achieved by reducing the radiation dose through adjustments in tube current and/or the number of projections.
View Article and Find Full Text PDFJ Neurophysiol
December 2024
Center for Neural Basis of Cognition, University of Pittsburgh and Carnegie Mellon University, Pittsburgh, PA, USA.
Surface electromyography () is useful for studying muscle function and controlling prosthetics, but crosstalk from nearby muscles often limits its effectiveness. High-density surface EMG () improves spatial resolution, allowing for the isolation of in the densely packed forearm muscles. This study assessed for localizing and evaluated the impact of spatial filters on crosstalk reduction.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!