Evaluation of interarm blood pressure differences using the Microlife WatchBP Office in a clinical setting.

Blood Press Monit

aDepartment of Endocrinology and Internal Medicine, Aarhus University Hospital bDepartment of Clinical Medicine cSection for Biostatistics, Department of Public Health, Aarhus University, Aarhus dDiagnostic Centre, Silkeborg Regional Hospital, Silkeborg eThe Danish Diabetes Academy, Odense University Hospital, Odense, Denmark.

Published: June 2017

Objective: The aim of this study was to evaluate the usefulness of Microlife WatchBP Office and the effect of increasing the number of measurements in the clinical evaluation of systolic interarm difference (IAD).

Patients And Methods: Office blood pressure was measured simultaneously on both arms in 339 patients (85% diabetic) using the Microlife WatchBP Office, a fully automatic, oscillometric device. The patients included were all scheduled for ambulatory blood pressure measurement at the outpatient clinic of endocrinology at Silkeborg Regional Hospital, Denmark. Two successive sets of three individual measurements were made. A statistical analysis of variance was carried out on the measurements.

Results: In the first set of measurements, the mean IAD was -0.3 mmHg and the prevalence of IAD greater than or equal to 10 mmHg was 9.1%. Only 7.6% of the patients with an IAD less than 10 mmHg in the first set of measurements had an IAD greater than or equal to 10 mmHg in the second set of measurements. The 95% limits of agreement for the mean IAD for a single set of three measurements were ±13.16 mmHg. The probability of detecting an IAD more than 10 mmHg only increased slightly with an increasing number of measurements.

Conclusion: A single set of triplicate measurements using Microlife WatchBP is an acceptable method for evaluating IAD as more measurements do little to improve the probability of detecting an IAD more than 10 mmHg because of high intraindividual variation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MBP.0000000000000246DOI Listing

Publication Analysis

Top Keywords

microlife watchbp
16
blood pressure
12
watchbp office
12
set measurements
12
iad 10 mmhg
12
increasing number
8
measurements
8
iad
8
measurements iad
8
iad greater
8

Similar Publications

Background: Early detection of atrial fibrillation (AF) is key for preventing strokes. Blood pressure monitors (BPMs) with built-in AF screening features have the potential for early detection at home. Recently, 2 BPMs (HEM-7371T1-AZ and HEM-7372T1-AZAZ, Omron Healthcare Co.

View Article and Find Full Text PDF

Introduction: Peripheral arterial disease is a marker of vascular damage that is diagnosed by measuring the ankle-brachial index. The aim of this study was to determine the validity and agreement of the MESI ABPI-MD and Microlife WatchBP® office-ABI oscillometric devices with respect to the gold standard arterial Doppler.

Materials And Methods: Observational, cross-sectional, descriptive study of inpatients who underwent ABI measurement with the three devices.

View Article and Find Full Text PDF

This study evaluated an oscillometric device (OD), Microlife WatchBP Office AFIB, and a hybrid manual auscultatory device (AD), Greenlight 300TM, to determine a suitable blood pressure (BP) measurement device for the Korea National Health and Nutrition Examination Survey in a mercury-free context. Adhering to the 2018 Universal Standard's suggested consensus, the study involved 800 subjects (mean age 51.2 ± 17.

View Article and Find Full Text PDF

The coexistence of hypertension and atrial fibrillation (AF) is common and accounts for a worse prognosis. Uncertainties exist regarding blood pressure (BP) measurements in AF patients by automated oscillometric devices. The Microlife WatchBP 03 AFIB ambulatory BP monitoring (ABPM) device including an AF algorithm with each measurement was used in 430 subjects aged >65 years referred for ABPM and with assumed paroxysmal AF to perform intra-individual comparisons of BP during both AF-indicated and sinus rhythm.

View Article and Find Full Text PDF

Accurate blood pressure (BP) measurement is crucial for hypertension detection and management. The Korea National Health and Nutrition Examination Survey (KNHANES) assesses the health of Koreans using representative cross-sectional data. BP measurements were historically done with mercury sphygmomanometers for participants aged ≥10 years.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!