Volume and Pressure Delivery During Pediatric High-Frequency Oscillatory Ventilation.

Pediatr Crit Care Med

1Department of Anesthesiology and Critical Care Medicine, Children's Hospital Los Angeles, Los Angeles, CA. 2Department of Pediatrics, Keck School of Medicine, University of Southern California, Los Angeles, CA. 3Department of Pediatric Critical Care, Geisinger Medical Center, Janet Weis Children's Hospital, Danville, PA.

Published: April 2017

Objective: Identify variables independently associated with delivered tidal volume (VT) and measured mean airway pressure during high-frequency oscillatory ventilation across the range of pediatric endotracheal tube sizes.

Design: In vitro study.

Setting: Research laboratory.

Interventions: An in vitro bench model of the intubated pediatric respiratory system during high-frequency oscillatory ventilation was used to obtain delivered VT and mean airway pressure (in the distal lung) for various endotracheal tube sizes. Measurements were taken at different combinations of ventilator set mean airway pressure (Paw), amplitude (ΔP), frequency, and test lung compliance. Multiple regression analysis was used to construct multivariable models predicting delivered VT and mean airway pressure.

Measurements And Main Results: Variables independently associated with higher delivered VT for all endotracheal tube sizes include higher ΔP (p < 0.001), lower frequency (p < 0.001), and higher test lung compliance (p < 0.001). A multiplicative interaction between frequency and ΔP magnifies the delivered VT when ΔP is high and frequency is low (p < 0.001). Delivered mean airway pressure becomes lower than set Paw as ΔP increases (p < 0.001) and frequency increases (p < 0.05). Ventilator set Paw is the largest determinant of delivered mean airway pressure; however, increasing ΔP resulted in a lower delivered mean airway pressure. For example, in a 4.0 mm ID endotracheal tube, increasing ΔP by 10 cm H2O resulted in an average decrease of delivered mean airway pressure by 4.5%.

Conclusions: This is the first study to quantify the interaction between ΔP and frequency in delivered VT and the effect of ΔP and frequency on delivered mean airway pressure. These results demonstrate the need to measure or estimate VT and delivered pressures during high-frequency oscillatory ventilation and may be useful in determining optimal strategies for lung protective ventilation during high-frequency oscillatory ventilation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/PCC.0000000000001089DOI Listing

Publication Analysis

Top Keywords

airway pressure
32
delivered airway
28
high-frequency oscillatory
20
oscillatory ventilation
20
endotracheal tube
16
delivered
12
Δp frequency
12
airway
9
Δp
9
variables independently
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!